Appendix C: Performance Verification
Select Start within the Sync frame at the bottom menu.
3.Turn on the AWG2021 CH1 output: If the LED above the CH1 output connector is not on, push the CH1 button.
4.Check autostep mode: Push the AWG2021 MANUAL TRIGGER button, and check that the oscilloscope momentarily displays one sine wave cycle with a different frequency and amplitude each time you push and release the button.
5.End procedure: Disconnect the oscilloscope.
Arithmetic Operation Checks
These procedures check operation of external AM. For an AWG2021 with
Option 02, they check internal AM and internal Add arithmetic functions.
NOTE. The arithmetic operation checks are structured as a continuous series of tests. After Check External AM Operation, each test uses the control settings from the last test and uses the next step in the sequence file.
Check External AM Operation
Electrical | External amplitude modulation, page)B*9. |
Characteristic Checked |
|
|
|
Equipment Required | Two 50)W coaxial cables, a 50 W terminator, a function |
| generator, and a digital multimeter DMM. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C*5. |
|
|
1.Install test hookup and set test equipment controls:
a.Hook up DMM: Connect the AWG2021 CH1 output through a coaxial cable, the 50 W terminator, and
b.Hook up function generator: Connect the AWG2021
C22 | AWG2021 User Manual |