Appendix C: Performance Verification
3.End procedure:
a.Disable function generator output: Turn the function generator output off.
b.Remove equipment: Disconnect connections to the test equipment.
Clock Frequency and Amplitude Checks
Check Clock Frequency Accuracy
These procedures check the accuracy of the AWG2021 clock frequency and the waveform output amplitude.
Electrical | Clock Generator, Accuracy, on page)B*6. |
Characteristic Checked |
|
|
|
Equipment Required | A 50)W coaxial cable, an SMA*BNC adapter, and a frequency |
| counter. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C*5. |
|
|
1.Install test hookup and set test equipment controls:
a.Hook up frequency counter: Connect the AWG2021 rear panel CLOCK OUT connector to the frequency counter input through a coaxial cable and an
Frequency Counter
DC 5010
AWG2021
AWG2021
Figure C#13: Clock Frequency Accuracy Initial Test Hookup