Appendix C: Performance Verification

3.End procedure:

a.Disable function generator output: Turn the function generator output off.

b.Remove equipment: Disconnect connections to the test equipment.

Clock Frequency and Amplitude Checks

Check Clock Frequency Accuracy

These procedures check the accuracy of the AWG2021 clock frequency and the waveform output amplitude.

Electrical

Clock Generator, Accuracy, on page)B*6.

Characteristic Checked

 

 

 

Equipment Required

A 50)W coaxial cable, an SMA*BNC adapter, and a frequency

 

counter.

 

 

Prerequisites

The AWG2021 meets the prerequisites listed on page C*5.

 

 

1.Install test hookup and set test equipment controls:

a.Hook up frequency counter: Connect the AWG2021 rear panel CLOCK OUT connector to the frequency counter input through a coaxial cable and an SMA-BNC adapter (see Figure C-13).

Frequency Counter

DC 5010

AWG2021

AWG2021

Figure C#13: Clock Frequency Accuracy Initial Test Hookup

AWG2021 User Manual