Appendix C: Performance Verification
b.Select the AWG2021 waveform file:
Push MODEAutostepSelect Autostep File to choose a sequence file for CH1.
Turn the general purpose knob to select the GAIN_OFF.AST file.
Push ENTER.
3.Check gain accuracy:
Check that the displayed step is Step 1 on the MODE menu. If it is not, select the side menu STOP button to return to Step 1.
Note the DMM reading as ªAº for this value.
Push the AWG2021 MANUAL TRIGGER button, and check that the displayed step is Step 2.
Note the DMM reading as ªBº for this value.
Do the following calculation.
A ± B
Check that the calculated value is in the range from 0.4925 V to 0.5075 V.
Push the AWG2021 MANUAL TRIGGER button, and check that the displayed step is Step 3.
Note the DMM reading as ªCº for this value.
Push the AWG2021 MANUAL TRIGGER button, and check that the displayed step is Step 4.
Record the DMM reading as ªDº for this value.
Do the following calculation.
C ± D
Check that the calculated value is in the range from 4.925 V to 5.075 V.
4.End procedure: Retain the test hookup and control settings.
C32 | AWG2021 User Manual |