Appendix C: Performance Verification
Press the numeric key 1, and press the units key V to select an amplitude of 1 V.
c.Repeat substeps 4a through e, checking to the follow limits:
Rise time . . . . . . . . . . . . . . . . . . . . . . . . 4.2 ns, maximum
Aberrations . . . . . . . . . . . . . . . . . . . . . . 0.4 div., maximum
Flatness . . . . . . . . . . . . . . . . . . . . . . . . . 0.15 div., maximum
Fall time . . . . . . . . . . . . . . . . . . . . . . . . 4.2 ns, maximum
6.Check pulse response for CH2 (Option 02): If the AWG2021 has a second channel, repeat this Pulse Response Check procedure using the AWG2021 CH2 output and selecting the waveform and setting controls for CH2.
7.End procedure: Remove the connections.
SYNC Out and MARKER Out Amplitude Checks
These procedures check the amplitude of the SYNC Out and MARKER Out signals.
Electrical | Auxlary Outputs, Sync, Marker 1, Marker 2, Ampltude, on |
Characteristic Checked | pae'B(7. |
|
|
Equipment Required | A 50'W coaxal cable and an osclloscope. |
|
|
Prerequisites | Te AWG2021 meets te prerequstes lsted on pae C(5. |
|
|
1.Install test hookup and set test equipment controls:
a.Hook up the oscilloscope: Connect the AWG2021 CH1 SYNC Out connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure
AWG2021 | Osclloscope |
AWG2021 | TDS 540 |
Figure C(17: Initial Test Hookup
C(36 | AWG2021 User Manual |