Appendix C: Performance Verification

￿Press the numeric key 1, and press the units key V to select an amplitude of 1 V.

c.Repeat substeps 4a through e, checking to the follow limits:

Rise time . . . . . . . . . . . . . . . . . . . . . . . . 4.2 ns, maximum

Aberrations . . . . . . . . . . . . . . . . . . . . . . 0.4 div., maximum

Flatness . . . . . . . . . . . . . . . . . . . . . . . . . 0.15 div., maximum

Fall time . . . . . . . . . . . . . . . . . . . . . . . . 4.2 ns, maximum

6.Check pulse response for CH2 (Option 02): If the AWG2021 has a second channel, repeat this Pulse Response Check procedure using the AWG2021 CH2 output and selecting the waveform and setting controls for CH2.

7.End procedure: Remove the connections.

SYNC Out and MARKER Out Amplitude Checks

These procedures check the amplitude of the SYNC Out and MARKER Out signals.

Electrical

Aux￿l￿ary Outputs, Sync, Marker 1, Marker 2, Ampl￿tude, on

Characteristic Checked

pa￿e'B(7.

 

 

Equipment Required

A 50'W coax￿al cable and an osc￿lloscope.

 

 

Prerequisites

T￿e AWG2021 meets t￿e prerequ￿s￿tes l￿sted on pa￿e C(5.

 

 

1.Install test hookup and set test equipment controls:

a.Hook up the oscilloscope: Connect the AWG2021 CH1 SYNC Out connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure C-17).

AWG2021

Osc￿lloscope

AWG2021

TDS 540

Figure C(17: Initial Test Hookup

C(36

AWG2021 User Manual