Appendix C: Performance Verification
Pulse Response Check
This procedure checks the pulse response characteristics of the AWG2021 output waveforms at amplitudes of 0.5 and 1 V.
Electrical | Main Outputs, Pulse Response, on page&B'7. |
Characteristic Checked |
|
|
|
Equipment Required | A 50&W coaxial cable and an oscilloscope. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C'5. |
|
|
1.Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure
AWG2021 | Oscilloscope |
AWG2021 | TDS 540 |
Figure C!16: Pulse Response Initial Test Hookup
b.Set oscilloscope controls:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . . DC
Scale . . . . . . . . . . . . . . . . . . . . . . . . 0.1 V/div.
Input impedance . . . . . . . . . . . . . . . 50 W
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . . . 5 ns/div.
Trigger
Source . . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . . Auto
C!34 | AWG2021 User Manual |