Appendix C: Performance Verification
Check Offset Accuracy
Electrical | Main Output, Offset, Accuracy, on page*B+6. |
Characteristic Checked |
|
|
|
Equipment Required | A 50*W coaxial cable, 50*W termination, BNC+to+dual banana |
| adapter, and a digital multimeter DMM. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C+5. |
|
|
1.Use the test hookup and test equipment settings from previous check, however, add a 50 W termination at the DMM input.
2.Check offset accuracy:
Push the AWG2021 MANUAL TRIGGER button, and check that the displayed step is Step 5.
Check that the DMM voltage reading is in the range from 2.470 V to 2.530 V.
Push the AWG2021 MANUAL TRIGGER button, and check that the displayed step is Step 6.
Check that the DMM voltage reading is in the range from ±0.005 V to 0.005 V.
Push the AWG2021 MANUAL TRIGGER button, and check that the displayed step is Step 7.
Check that the DMM voltage reading is in the range from ±2.470 V to ±2.530 V.
3.Check Option 02: If the AWG2021 has Option 02, repeat the Gain Accuracy and Offset Accuracy Checks for the AWG2021 channel 2 (CH2).
4.End procedure: Disconnect the DMM.
AWG2021 User Manual | C33 |