Appendix C: Performance Verification
4.Check
Remove the coaxial cable from the AWG2021
Check that the pulse amplitude of the displayed waveform is from 2.250
5.Check Option 02: If the AWG2021 has a second channel, repeat this entire test, selecting the AWG2021 waveform and setting controls for CH2 and checking:
6.End procedure: Disconnect the oscilloscope.
External Trigger Level Accuracy Check
This procedure checks the external trigger level accuracy of the AWG2021.
Electrical | Auxiliary Inputs, Trigger, Accuracy, on page&B'9. |
Characteristic Checked |
|
|
|
Equipment Required | Two 50&W coaxial cables, a function generator, and an |
| oscilloscope. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C'5. |
|
|
1.Install test hookup and set test equipment controls:
a.Hook up oscilloscope: Connect the AWG2021 CH1 output through a coaxial cable to the oscilloscope CH1 vertical input.
b.Hook up function generator: Connect the AWG2021 TRIGGER INPUT through a coaxial cable to the function generator output (see Fig-
ure