Appendix C: Performance Verification
b.Change the AWG2021 controls:
Push MODEPolarity to highlight Negative.
Select Level from the side menu, and turn the general purpose knob to select ±1 V. (You can also use the numeric and units keys to select ±1 V; then push ENTER.)
c.Check external trigger level accuracy:
Gradually decrement the function generator offset level until a waveform is displayed on the oscilloscope.
Check that that the function generator offset level is from ±1.15 V to ±0.85 V, when the waveform is first displayed.
6.End procedure: Turn off the function generator output and disconnect the function generator.
External CLOCK IN Check
This procedure checks the AWG2021 response to an external CLOCK IN signal.
Electrical | Auxiiary Iputs, Cck, Threshd eve, page(B)9. |
Characteristic Checked |
|
|
|
Equipment Required | Tw 50(W caxia cabes, a fucti geeratr, ad a |
| sciscpe. |
|
|
Prerequisites | The AWG2021 eets the prerequisites isted page C)5. |
|
|
1.Install test hookup and set test equipment controls:
a.Hook up oscilloscope: Connect the AWG2021 CH1 output through a coaxial cable to the oscilloscope CH1 vertical input.
b.Hook up function generator: Connect the AWG2021