Appendix C: Performance Verification
2.Select the AWG2021 waveform file and set AWG2021 controls:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the EXT_CLK.WFM file.
Push ENTER.
3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check the external CLOCK IN threshold level:
a.Enable function generator output: Turn on function generator output.
b.Check the level: Check that the waveform displayed on the oscilloscope has an amplitude of 5 divisions and a stable display of 5 cycles.
5.Turn off equipment output and disconnect test hookup:
a.Disable function generator output: Turn off function generator output.
b.Remove connections: Disconnect all connections to the AWG2021.
ECL Digital Data Out Check (Option 03)
This procedure checks the AWG2021 ECL digital data output at the rear panel.
This check requires that the AWG2021 has Option 03 installed.
Electrical | Auxiliary Outputs, ECL Digital Data Out, Level, on page)B*8. |
Characteristic Checked |
|
|
|
Equipment Required | Test leads, ECL digital data out cable, power supply, |
| termination board, and oscilloscope. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C*5. |
|
|
AWG2021 User Manual | C 43 |