Appendix C: Performance Verification
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . . . Adjust as needed
Trigger
Mode . . . . . . . . . . . . . . . . . . . . . . . Auto
e.Set power supply controls:
Parameter
Supply select . . . . . . . . . . . . . . . . . Negative Voltage . . . . . . . . . . . . . . . . . . . . . . 2
2.Select the AWG2021 waveform file, and set AWG2021 controls:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1.
Turn the general purpose knob to highlight the DIGI_OUT.WFM file.
Push ENTER to select the file.
c.Set AWG2021 controls:
Push MODECont
3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check the digital data output signals:
a.Enable power supply output: Turn on power supply output.
b.Check the signal levels:
Contact the oscilloscope probe to the pins on J200 and J210 (see Figure
Data signals D0±D11, D0±D11 are differential ECL output. Clock signals CLK and CLK are differential ECL output. All other pins are ground.
AWG2021 User Manual | C45 |