Appendix C: Performance Verification

b.Set oscilloscope controls:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

Coupling . . . . . . . . . . . . . . . . . . . . . DC

Scale . . . . . . . . . . . . . . . . . . . . . . . . 500 mV/div.

Input impedance . . . . . . . . . . . . . . . 50 W

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . . . 500 ns/div.

Trigger

Source . . . . . . . . . . . . . . . . . . . . . . CH1

Coupling . . . . . . . . . . . . . . . . . . . . . DC

Slope . . . . . . . . . . . . . . . . . . . . . . . Positive

Level . . . . . . . . . . . . . . . . . . . . . . . 500 mV

Mode . . . . . . . . . . . . . . . . . . . . . . . Auto

2.Set the AWG2021 controls and select the waveform file:

a.Initialize AWG2021 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Select the waveform file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to select the CLK_AMPL.WFM file.

￿Push ENTER.

3.Check clock amplitude accuracy: Check that the pulse displayed on the oscilloscope has an amplitude of 0.7 Vp-pto 1.3 Vp-p.

4.End procedure: Disconnect the oscilloscope.