Appendix C: Performance Verification
Table C%3: Clock Frequency Accuracy
Clock Frequency | Frequency Range |
|
|
150 MHz | 149.9925.MHz - 150.0075.MHz |
|
|
1 MHz | 0.99995.MHz ± 1.00005.MHz |
|
|
1 kHz | 0.99995.kHz - 1.00005.kHz |
|
|
10 Hz | 9.9995.Hz - 10.0005.Hz |
|
|
4.End procedure: Disconnect the frequency counter.
Check Clock Amplitude
Electrical | Auxiliary Outputs, Clock, Amplitude, on page.B/8. |
Characteristic Checked |
|
|
|
Equipment Required | A 50.W coaxial cable, an SMA/BNC adapter, and an |
| oscilloscope. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C/5. |
|
|
1.Install test hookup and set test equipment controls:
a.Hook up oscilloscope: Connect the AWG2021
Oscilloscope
TDS 540
AWG2021
AWG2021
Figure C%14: Clock Amplitude Initial Test Hookup
AWG2021 User Manual | C%29 |