Appendix C: Performance Verification

Table C%3: Clock Frequency Accuracy

Clock Frequency

Frequency Range

 

 

150 MHz

149.9925.MHz - 150.0075.MHz

 

 

1 MHz

0.99995.MHz ± 1.00005.MHz

 

 

1 kHz

0.99995.kHz - 1.00005.kHz

 

 

10 Hz

9.9995.Hz - 10.0005.Hz

 

 

4.End procedure: Disconnect the frequency counter.

Check Clock Amplitude

Electrical

Auxiliary Outputs, Clock, Amplitude, on page.B/8.

Characteristic Checked

 

 

 

Equipment Required

A 50.W coaxial cable, an SMA/BNC adapter, and an

 

oscilloscope.

 

 

Prerequisites

The AWG2021 meets the prerequisites listed on page C/5.

 

 

1.Install test hookup and set test equipment controls:

a.Hook up oscilloscope: Connect the AWG2021 rear-panel CLOCK OUT connector through a coaxial cable and SMA-BNC adapter to the oscilloscope CH1 vertical input (see Figure C-14).

Oscilloscope

TDS 540

AWG2021

AWG2021

Figure C%14: Clock Amplitude Initial Test Hookup

AWG2021 User Manual

C%29