Appendix C: Performance Verification

Check Internal AM Operation (Option 02 only)

Electrical

Arithmetic Operation, Amplitude Modulation, on page+B,9.

Characteristic Checked

 

 

 

Equipment Required

Two 50+W coaxial cables, a function generator, and a digital

 

multimeter ￿DMM￿.

 

 

Prerequisites

The AWG2021 meets the prerequisites listed on page C,5. It

 

also requires Option 02 ￿CH2 output channel￿ in the AWG2021.

 

 

1.Use test hookup and control settings from previous check.

2.Check internal AM operation:

a.Check Autostep Step 2:

￿Push the AWG2021 MANUAL TRIGGER button, and check that the step changes to Step 2 on the MODE menu.

￿Check that the DMM reading is in the range from 2.375 VDC to 2.625 VDC.

b.Check Autostep Step 3:

￿Push the AWG2021 MANUAL TRIGGER button, and check that the Autostep changes to Step: 3 on the MODE menu.

￿Check that the DMM reading is in the range from ±2.625 V to ±2.375 V.

3.End procedure: Retain the test hookup and settings for the next check.

Check Internal Add Operation (Option 02 only)

Electrical

Arithmetic Operation, Add, on page+B,9.

Characteristic Checked

 

 

 

Equipment Required

Two 50+W coaxial cables, a function generator, and a digital

 

multimeter ￿DMM￿.

 

 

Prerequisites

The AWG2021 meets the prerequisites listed on page C,5. It

 

also requires Option 02 ￿CH2 output channel￿ in the AWG2021.

 

 

1.Use test hookup and control settings from previous check.

2.Check internal Add operation: a. Check Autostep Step 4:

￿Push the AWG2021 MANUAL TRIGGER button, and check that the step changes to Step 4 on the MODE menu.

￿Check that the DMM reading is in the range from 2.85 V to 3.15 V.

C!26

AWG2021 User Manual