Appendix C: Performance Verification
Check Internal AM Operation (Option 02 only)
Electrical | Arithmetic Operation, Amplitude Modulation, on page+B,9. |
Characteristic Checked |
|
|
|
Equipment Required | Two 50+W coaxial cables, a function generator, and a digital |
| multimeter DMM. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C,5. It |
| also requires Option 02 CH2 output channel in the AWG2021. |
|
|
1.Use test hookup and control settings from previous check.
2.Check internal AM operation:
a.Check Autostep Step 2:
Push the AWG2021 MANUAL TRIGGER button, and check that the step changes to Step 2 on the MODE menu.
Check that the DMM reading is in the range from 2.375 VDC to 2.625 VDC.
b.Check Autostep Step 3:
Push the AWG2021 MANUAL TRIGGER button, and check that the Autostep changes to Step: 3 on the MODE menu.
Check that the DMM reading is in the range from ±2.625 V to ±2.375 V.
3.End procedure: Retain the test hookup and settings for the next check.
Check Internal Add Operation (Option 02 only)
Electrical | Arithmetic Operation, Add, on page+B,9. |
Characteristic Checked |
|
|
|
Equipment Required | Two 50+W coaxial cables, a function generator, and a digital |
| multimeter DMM. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C,5. It |
| also requires Option 02 CH2 output channel in the AWG2021. |
|
|
1.Use test hookup and control settings from previous check.
2.Check internal Add operation: a. Check Autostep Step 4:
Push the AWG2021 MANUAL TRIGGER button, and check that the step changes to Step 4 on the MODE menu.
Check that the DMM reading is in the range from 2.85 V to 3.15 V.
C!26 | AWG2021 User Manual |