Appendix C: Performance Verification
Turn the general purpose knob to highlight the OPE.AST file.
Push ENTER.
3.Turn on the AWG2021 CH1 output: If the LED above the CH1 output connector is not on, push the CH1 button.
4.Enable the function generator output: Turn on the function generator output.
5.Check external AM operation:
Check that the step number displayed on the AWG2021 MODE menu is Step 1 (see Figure
Step Display
Figure C12: MODE Menu Autostep Setting
Check that the DMM reading is in the range from 2.375 to 2.625 V (100% modulation).
Set the function generator offset value to 0 V. Check that the DMM reading is in the range from 1.125 to 1.375 V (50% modulation).
C24 | AWG2021 User Manual |