Appendix C: Performance Verification

￿Turn the general purpose knob to highlight the OPE.AST file.

￿Push ENTER.

3.Turn on the AWG2021 CH1 output: If the LED above the CH1 output connector is not on, push the CH1 button.

4.Enable the function generator output: Turn on the function generator output.

5.Check external AM operation:

￿Check that the step number displayed on the AWG2021 MODE menu is Step 1 (see Figure C-12). If it is not, push MANUAL TRIGGER to step though the sequence file steps until Step: 1 is displayed.

Step Display

Figure C￿12: MODE Menu Autostep Setting

￿Check that the DMM reading is in the range from 2.375 to 2.625 V (100% modulation).

￿Set the function generator offset value to 0 V. Check that the DMM reading is in the range from 1.125 to 1.375 V (50% modulation).

C￿24

AWG2021 User Manual