Appendix C: Performance Verification

Function Generator

FG 5010

AWG2021

AWG2021

DMM

8842A

Figure C￿11: External AM Operation Initial Test Hookup

c.Set DMM controls:

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC

Range . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . Front

d.Set function generator controls:

Function . . . . . . . . . . . . . . . . . . . . . . . . Square

Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous

Parameter

Frequency . . . . . . . . . . . . . . . . . . . . 1 kHz

Amplitude . . . . . . . . . . . . . . . . . . . 0 V

Offset . . . . . . . . . . . . . . . . . . . . . . . 1 V

Output . . . . . . . . . . . . . . . . . . . . . . . . . . Off

2.Set the AWG2021 controls and select the waveform file:

a.Initialize AWG2021 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Modify AWG2021 default settings:

￿Push MODE￿Autostep.

c.Select waveform file:

￿Select Select Autostep File from the side menu to choose a waveform file for CH1.

AWG2021 User Manual

C￿23