Appendix C: Performance Verification
Function Generator
FG 5010
AWG2021
AWG2021
DMM
8842A
Figure C11: External AM Operation Initial Test Hookup
c.Set DMM controls:
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . Front
d.Set function generator controls:
Function . . . . . . . . . . . . . . . . . . . . . . . . Square
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Parameter
Frequency . . . . . . . . . . . . . . . . . . . . 1 kHz
Amplitude . . . . . . . . . . . . . . . . . . . 0 V
Offset . . . . . . . . . . . . . . . . . . . . . . . 1 V
Output . . . . . . . . . . . . . . . . . . . . . . . . . . Off
2.Set the AWG2021 controls and select the waveform file:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Modify AWG2021 default settings:
Push MODEAutostep.
c.Select waveform file:
Select Select Autostep File from the side menu to choose a waveform file for CH1.
AWG2021 User Manual | C23 |