
Appendix C: Performance Verification
WG2021 | Oscilloscope |
AWG2021 | TDS 540 |
Figure C10: Autostep Mode Initial Test Hookup
b.Set the oscilloscope controls:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1
CH1 coupling . . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . . . 0.5 V/div.
CH1 input impedance . . . . . . . . . . 50 W
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . . . 2 ms/div.
Trigger
Source . . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . . 100 mV
Mode . . . . . . . . . . . . . . . . . . . . . . . Normal
2.Set the AWG2021 controls and select the waveform file:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Modify AWG2021 default settings:
Push MODEAutostepSlope to highlight Positive.
Select Level from the side menu, and turn the general purpose knob to select 1 V.
Select Select Autostep File from the side menu to choose from the file list for CH1.
Turn the general purpose knob to highlight the MODE_AST.AST file.
Push ENTER.
AWG2021 User Manual | C21 |