Appendix C: Performance Verification

Check Waveform Advance Mode

Electrical

￿pe￿a ing mode, Wa"efo￿m Ad"ance, on page$B%1.

Characteristic Checked

 

 

 

Equipment Required

A 50$W coaxial cable and an o￿cillo￿cope.

 

 

Prerequisites

The AWG2021 mee ￿ he p￿e￿eq!i￿i e￿ li￿ ed on page C%5.

 

 

1.Install test hookup and set test equipment controls:

a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure C-9).

AWG2021

￿￿cillo￿cope

AWG2021

TDS 540

Figure C$9: Waveform Advance Mode Initial Test Hookup

b.Set oscilloscope controls:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling . . . . . . . . . . . . . . . . . DC

CH1 . . . . . . . . . . . . . . . . . . . . . . . . 0.2 V/div.

CH1 input impedance . . . . . . . . . . 50 W

Horizontal . . . . . . . . . . . . . . . . . . . . . . . .

Sweep . . . . . . . . . . . . . . . . . . . . . . . 5 ms/div.

Trigger

Source . . . . . . . . . . . . . . . . . . . . . . CH1

Coupling . . . . . . . . . . . . . . . . . . . . . DC

Slope . . . . . . . . . . . . . . . . . . . . . . . Positive

Level . . . . . . . . . . . . . . . . . . . . . . . 0 V

Mode . . . . . . . . . . . . . . . . . . . . . . . Auto

2.Set the AWG2021 controls and select the waveform file:

a.Initialize AWG2021 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Set AWG2021 controls:

AWG2021 User Manual

C$19