Appendix C: Performance Verification
Check Waveform Advance Mode
Electrical | pea ing mode, Wa"efom Ad"ance, on page$B%1. |
Characteristic Checked |
|
|
|
Equipment Required | A 50$W coaxial cable and an ocillocope. |
|
|
Prerequisites | The AWG2021 mee he peeq!ii e li ed on page C%5. |
|
|
1.Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure
AWG2021 | cillocope |
AWG2021 | TDS 540 |
Figure C$9: Waveform Advance Mode Initial Test Hookup
b.Set oscilloscope controls:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1
CH1 coupling . . . . . . . . . . . . . . . . . DC
CH1 . . . . . . . . . . . . . . . . . . . . . . . . 0.2 V/div.
CH1 input impedance . . . . . . . . . . 50 W
Horizontal . . . . . . . . . . . . . . . . . . . . . . . .
Sweep . . . . . . . . . . . . . . . . . . . . . . . 5 ms/div.
Trigger
Source . . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . . Auto
2.Set the AWG2021 controls and select the waveform file:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Set AWG2021 controls:
AWG2021 User Manual | C$19 |