Appendix C: Performance Verification
c.Select the file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the MODE.WFM file.
Push ENTER to select the file.
3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check burst count: Push the AWG2021 MANUAL TRIGGER button and check that when the button is pushed, the oscilloscope displays three cycles of sine wave.
5.End procedure: Disconnect the oscilloscope.
Check Gated Mode
Electrical | pea ing mode, Ga ed, on page#B$1. |
Characteristic Checked |
|
|
|
Equipment Required | Thee 50#W coaxial cable, a 50 W peciion emina ion, a |
| f!nc ion genea o, and an ocillocope. |
|
|
Prerequisites | The AWG2021 mee he peeq!ii e li ed on page C$5. |
|
|
1.Install test hookup and set test equipment controls:
a.Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope.
b.Hook up the function generator: Connect the function generator output to both the AWG2021 TRIGGER INPUT and the oscilloscope CH2 input through a coaxial cable, precision termination, and a dual input coupler (see Figure
AWG2021 User Manual | C15 |