Appendix C: Performance Verification

2.Set AWG2021 controls and select the waveform file:

a.Initialize AWG2021 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Modify the AWG2021 default settings:

￿Push MODE￿Triggered￿Slope to select Positive slope.

￿Select Level from the side menu and turn the general purpose knob to select a 1 V trigger level.

￿Select Impedance from the side menu to select 50 W impedance.

c.Select the file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Highlight the MODE.WFM file using the general purpose knob.

￿Push ENTER to select the file.

3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check triggered mode with manual triggering: Push the AWG2021 MANUAL TRIGGER button and check that when the button is pushed, the oscilloscope displays a one-cycle sine wave.

5.Check triggered mode with external triggering:

a.Enable function generator output: Turn on the function generator output.

b.Check triggering: Check that for each trigger supplied by the function generator, the oscilloscope displays a one-cycle sine wave.

6.End procedure: Turn off the function generator output, and disconnect the function generator and oscilloscope.