Appendix C: Performance Verification
2.Set AWG2021 controls and select the waveform file:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Modify the AWG2021 default settings:
Push MODETriggeredSlope to select Positive slope.
Select Level from the side menu and turn the general purpose knob to select a 1 V trigger level.
Select Impedance from the side menu to select 50 W impedance.
c.Select the file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE.WFM file using the general purpose knob.
Push ENTER to select the file.
3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check triggered mode with manual triggering: Push the AWG2021 MANUAL TRIGGER button and check that when the button is pushed, the oscilloscope displays a
5.Check triggered mode with external triggering:
a.Enable function generator output: Turn on the function generator output.
b.Check triggering: Check that for each trigger supplied by the function generator, the oscilloscope displays a
6.End procedure: Turn off the function generator output, and disconnect the function generator and oscilloscope.
AWG2021 User Manual | C13 |