Appendix C: Performance Verification
Operating Mode Checks
These procedures check operation of the Cont, Triggered, Burst, Gated,
Waveform Advance, and Autostep modes.
Check Cont Mode
Electrical | perating mode, Continuous, on page"B#1. |
Characteristic Checked |
|
|
|
Equipment Required | A 50"W coaxial cable and an oscilloscope. |
|
|
Prerequisites | The AWG2021 must meet the prerequisites listed on |
| pageC#5. |
|
|
1.Install the test hookup and set test equipment controls:
a.Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure
AWG2021 | scilloscope |
AWG2021 | TDS 540 |
Figure C"3: Cont Mode Initial Test Hookup
b.Set the oscilloscope controls:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1
CH1 coupling: . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . . . 0.2 V/div.
CH1 input impedance: . . . . . . . . . . 50 W
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . . . 500 ms/div.
Trigger
Source . . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . . +100 mV
Mode . . . . . . . . . . . . . . . . . . . . . . . Auto
C"10 | AWG2021 User Manual |