Appendix C: Performance Verification

Operating Mode Checks

These procedures check operation of the Cont, Triggered, Burst, Gated,

Waveform Advance, and Autostep modes.

Check Cont Mode

Electrical

￿perating mode, Continuous, on page"B#1.

Characteristic Checked

 

 

 

Equipment Required

A 50"W coaxial cable and an oscilloscope.

 

 

Prerequisites

The AWG2021 must meet the prerequisites listed on

 

page￿C#5.

 

 

1.Install the test hookup and set test equipment controls:

a.Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure C-3).

AWG2021

￿scilloscope

AWG2021

TDS 540

Figure C"3: Cont Mode Initial Test Hookup

b.Set the oscilloscope controls:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1

CH1 coupling: . . . . . . . . . . . . . . . . DC

CH1 scale . . . . . . . . . . . . . . . . . . . . 0.2 V/div.

CH1 input impedance: . . . . . . . . . . 50 W

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . . . 500 ms/div.

Trigger

Source . . . . . . . . . . . . . . . . . . . . . . CH1

Coupling . . . . . . . . . . . . . . . . . . . . . DC

Slope . . . . . . . . . . . . . . . . . . . . . . . Positive

Level . . . . . . . . . . . . . . . . . . . . . . . +100 mV

Mode . . . . . . . . . . . . . . . . . . . . . . . Auto

C"10

AWG2021 User Manual