Appendix C: Performance Verification

memory. For instructions on loading files, see Loading Sample Waveforms on page 2 -28 in section 2.

Related Information Read Preparation and Conventions on page C-1. Also, if you are not familiar with operating the AWG2021, read in section 2 before doing any of these procedures.

Equipment Required The following equipment is required to check the performance of the AWG2021.

Table C￿1: Test Equipment

Item

Minimum Requirements

Example

Purpose

 

 

 

 

Precision termination

Impedance: 50?W, 0.1%

Tektronix Part 011@0129@XX

Signal termination.

 

Connectors: BNC

 

 

 

 

 

 

Adapter

Connectors: BNC female@to@dual

Tektronix Part 103@0090@XX

Signal interconnection.

 

banana

 

 

 

 

 

 

Adapter

Connectors: SMA male@to@BNC

Tektronix Part 015@0554@XX

Signal interconnection.

 

female

 

 

 

 

 

 

BNC dual input (TEE) adapter

Connectors: BNC

Tektronix Part 103@0030@XX

Signal interconnection.

 

 

 

 

BNC cable (4 required)

Impedance 50?W

Tektronix Part 012@0057@XX

Signal interconnection.

 

Connectors: BNC

 

 

 

Length: 43 inches

 

 

Termination board (Option 03)

Must use example equipment

Tektronix Part 671@2957@XX

Used to check ECL digital

 

 

 

data output.

 

 

 

 

ECL Digital Data Out Cable

Must use example equipment

Tektronix Part 012@1408@XX

Used to check ECL digital

(Option 03)

 

(Standard accessory for

data output.

 

 

Option 03)

 

 

 

 

 

Test leads (Option 03)

Must use example equipment

Tektronix Part 012@1381@XX

Used to check ECL digital

 

 

 

data output.

 

 

 

 

TTL Digital Data Out Cable

Must use example equipment

Tektronix Part 174@3192@XX

Used to check TTL digital

(Option 04)

 

(Standard accessory for

data output.

 

 

Option 04)

 

 

 

 

 

2 13 header (Option 04)

Must use example equipment

Tektronix Part 131@3847@XX

Used to check TTL digital

 

 

 

data output.

 

 

 

 

Probe, 10X (Option 04)

10X probe

Tektronix Part P6139A

Used to check TTL digital

 

 

 

data output.

 

 

 

 

Test oscilloscope

Bandwidth: >250?MHz

Tektronix TDS500 Series

Checks output signals. Used

 

 

Digitizing Oscilloscope or

in many procedures.

 

 

2400 Series Digitizing Oscillo@

 

 

 

scope

 

 

 

 

 

Frequency counter

Frequency range: 10?Hz to 250?MHz

Tektronix DC 5010 Program@

Used to check clock frequen@

 

 

mable Universal Counter/Tim@

cy.

 

 

er*

 

C￿6

AWG2021 User Manual