Teledyne Reference Gas, Zero Gas, Span Gas, Thermal Conductivity Analyzer, Model 212R

Models: 212R

1 28
Download 28 pages 9.87 Kb
Page 10
Image 10
2.3.1 Reference Gas.

Thermal Conductivity Analyzer

Model 212R

 

 

 

 

cause a large fluctuation in output signal. For best results, use electropolished SS tubing

Note #4: Becauseofthediffusionofairthroughcompositionmaterials, only dual stage SS regulators with metallic diaphragms should be used in conjunction with the sample and supporting gas

supplies.

Note #5: Regulators must be thoroughly purged (burped) prior to use.

2.3.1 Reference Gas.

A cylinder of gas of fixed composition is required as the reference for comparison with the sample gas. A total lack of impurities in the reference would be ideal but is not necessary. An impurity concentration of up to 50% of the narrowest range of interest is tolerable. The exact composition of the reference gas is academic as long as the user is certain that it falls within the limits specified in Section 7 of the manual. The important consideration is that the compositionofthereferencegasremainunchangedwheninuse.

NOTE: When it becomes necessary to replace the reference gas, the analyzer will have to be recalibrated.

2.3.2 Zero Gas.

A supply of gas, composed of the background gas of the analysis and the lowest attainable concentration of impurity, will be required to standard- ize one end of the ranges of interest. The composition of the zero gas must be known to the same exactitude expected of the analysis. TAI suggests that the gas be obtained from a supplier that will certify its composition. Recom- mendations as to the composition of the gas will be found in Section 7 of the manual.

2.3.3 Span Gas.

A supply of gas, composed of the background of the analysis plus a known concentration of the component of interest, will be required to stan- dardize the sensitivity of the analyzer. Again, the composition of the gas must be known to the same order of accuracy expected of the analyzer. Analysis and certification of the composition is desirable. Specific recom- mendations governing the composition of the span gas will be found in Section 7 of the manual.

10

Teledyne Analytical Instruments

Page 10
Image 10
Teledyne Reference Gas, Zero Gas, Span Gas, Thermal Conductivity Analyzer, Model 212R, Teledyne Analytical Instruments