Thermal Conductivity Analyzer | Model 212R |
|
|
|
|
CAUTION: Do not touch the thermistor probe with the iron. The end of the probe is tied and taped to the sample tubing below the cell cover. If the meter reading suddenly drops to zero, remove the soldering iron immediately. After the area around the thermistor cools, the voltage across the heater resistors should again climb to 47 volts. Such action indicates a properly functioning heater circuit. If the meter stays in the 47 volt region, the circuit card will have to be replaced.
7.0CALIBRATION DATA FOR MODEL 212R SERIAL NUMBER: 195929
7.1Ranges.
The ranges of the analyzer are:
RangeSwitchPosition#1: |
|
|
|
Range Switch Position #2: |
|
|
|
Range Switch Position #3: |
|
|
|