Thermal Conductivity Analyzer | Model 212R |
|
|
|
|
APPENDIX
Specifications:
Ranges: Three decade ranges: Minimum Full scale(ppm) 20 ppm Hydrogen balance Argon
25 ppm Hydrogen balance Nitrogen, Air
100 ppm Nitrogen balance Argon, Helium
100ppmHeliumbalanceAir
150 ppm Nitrogen, Carbon Monoxide balance Hydrogen
200 ppm Oxygen balance Argon
Range ID: 3 range ID contacts
Local Range ID LED’S
Detector: Thermal conductivity sensor
Signal Output:
Accuracy: ±2%fullscaleformost binarymixture at constant temperature. ± 5% of Full Scale over operating temperature (once temperatureequilibriumhasbeen achieved)
Operating
Temperature: 20 – 30°C (68° F to 85° F)
Sample
Requirement: Sample:0.3SCFH
Reference:0.1SCFH
Display: DigitalLEDreadout(3½Digit)
Mounting: Flushpanelmount.
Response Time: 90% of full scale less then 60 seconds.
Formostapplication.
Stability: Less then 2% full scale drift over 24 hours
Power
requirement: 110VAC, 50/60Hz (220VAC optional)
28 | Teledyne Analytical Instruments |