Teledyne Thermal Conductivity Analyzer, Model 212R, APPENDIX Specifications, Operating, Sample

Models: 212R

1 28
Download 28 pages 9.87 Kb
Page 28
Image 28
APPENDIX

Thermal Conductivity Analyzer

Model 212R

 

 

 

 

APPENDIX

Specifications:

Ranges: Three decade ranges: Minimum Full scale(ppm) 20 ppm Hydrogen balance Argon

25 ppm Hydrogen balance Nitrogen, Air

100 ppm Nitrogen balance Argon, Helium

100ppmHeliumbalanceAir

150 ppm Nitrogen, Carbon Monoxide balance Hydrogen

200 ppm Oxygen balance Argon

Range ID: 3 range ID contacts

Local Range ID LED’S

Detector: Thermal conductivity sensor

Signal Output: 0-1VDC negative ground 4-20maDC Isolated ground

Accuracy: ±2%fullscaleformost binarymixture at constant temperature. ± 5% of Full Scale over operating temperature (once temperatureequilibriumhasbeen achieved)

Operating

Temperature: 20 – 30°C (68° F to 85° F)

Sample

Requirement: Sample:0.3SCFH

Reference:0.1SCFH

Display: DigitalLEDreadout(3½Digit)

Mounting: Flushpanelmount.

Response Time: 90% of full scale less then 60 seconds.

Formostapplication.

Stability: Less then 2% full scale drift over 24 hours

Power

requirement: 110VAC, 50/60Hz (220VAC optional)

28

Teledyne Analytical Instruments

Page 28
Image 28
Teledyne instruction manual Thermal Conductivity Analyzer, Model 212R, APPENDIX Specifications, Operating, Sample, Power