TAPI Model 265A Chemiluminescence Ozone Analyzer
Manual Addendum, Rev. B, DCN 5254
14
PRINTED DOCUMENTS ARE UNCONTROLLED
Table 2-2: Final Test and Calibration Values (Continued)
Parameter Observed
Value Units Nominal Range Reference Section
O3 Span Conc ppb 0.5 - 5000
O3 Slope 1.0 ± 0.3
O3 Offset mV ± 25
Noise at Zero ppb 0.0 – 1.0
Noise At Span ppb 0.1 – 7.0
Measured Flows
Sample Flow cm³/min 500 ± 50
Factory Installed Options Option Installed
Power Voltage/Frequency
PROM # Serial #
Date Technician