TAPI Model 265A Chemiluminescence Ozone Analyzer
Manual Addendum, Rev. B, DCN 5254
23
PRINTED DOCUMENTS ARE UNCONTROLLED
5.1 Maintenance Schedule
NOTE
The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Table 5-1: Preventative Maintenance Schedule
Item Maintenance Interval Reference Section
TEST functions Review monthly See M200A manual, Table 8-1
Zero/Span calibration Annually or after repairs See M200A manual, Table 8-1
Zero/Span checks Daily See M200A manual, Table 8-1
Particulate filter Weekly as needed See M200A manual, Table 8-1
Reaction cell window Clean annually or as necessary See M200A manual, Table 8-1
Nitric oxide flow Check TEST function every
month
Section 1-3
See also M200A manual, Table
8-1 (O3 flows)
Sample Flow Check TEST function every
month
See M200A manual, Table 8-1
Pneumatic lines Examine every 12 months,
clean if necessary.
See M200A manual, Table 8-1
Factory calibration Calibrate each year or after
repairs
See M200A manual, Table 8-1
Leak Check Check every year or after other
maintenance.
See M200A manual, Table 8-1
Reaction cell O-rings Replace every 12 months See M200A manual, Table 8-1
Other O-rings Replace every 12 months See M200A manual, Table 8-1