3 Installation Model 3010MB
3-6: Part I Teledyne Analytical Instruments
Table 3-1: Analog Concentration Output-Examples
Voltage Signal Current Signal
%O2 Output (V dc) Output (mA dc)
0 0.0 4.0
1 0.1 5.6
2 0.2 7.2
3 0.3 8.8
4 0.4 10.4
5 0.5 12.0
6 0.6 13.6
7 0.7 15.2
8 0.8 16.8
9 0.9 18.4
10 1.0 20.0
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.
Table 3-2: Analog Range ID Output - Example
Range Voltage (V) Current (mA)
LO 0.25 8
MED 0.50 12
HI 0.75 16
CAL (0-25%) 1.00 20