5 Maintenance Model 3010MB Percent Paramagnetic Oxygen Analyzer
5-2: Part I Teledyne Analytical Instruments
Figure 5-1: Removing Fuse Block Cap and Fuse from Housing
2. Replace fuse by reversing process in step 1.
5.2 System Self Diagnostic Test
1. Press the
System
button to enter the system mode.
2. Use the < > arrow keys to move to More, and press
Enter
.
3. Use the < > arrow keys to move to Self-Test, and press
Enter
.
The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0OK
1 5 V Failure
2 15 V Failures
3 Both Failed
Analog
0OK
1 DAC A (0–1 V Concentration)
2 DAC B (0–1 V Range ID)
3 Both Failed