06712B (DCN5875) 3-1
3 GETTING STARTED
This section provides instructions on the proper installation and power up of the M701H.
Refer to Figure 3-1 for the instrument layout and Table 2-1 for the specifications.

3.1 Unpacking

CAUTION
TO AVOID PERSONAL INJURY, ALWAYS USE TWO PERSONS TO
LIFT AND CARRY THE M701H.
1. Verify that there is no shipping damage. If there are signs of damage, immediately
advise the shipper, then Teledyne API.
2. Remove the M701H from its shipping carton.
3. Remove the cover and check for damage inside.
4. Remove the four red shipping screws that secure the compressor base to the bottom
of the chassis. Refer to Figure 5-2 for the location of the screws.
5. Check the line voltage and frequency label on the rear panel to ensure that it
matches the local power source.

3.2 Installation

The basic M701H has rubber feet for counter-top use. Optionally, the M701H is supplied
with slides and angles for mounting in a standard 19” RETMA rack.
In order to ensure safe operation of the M701H, the following steps must be taken prior
to installation:
1. At least 6” (150 mm) clearance at the back for proper fan operation.
2. At least 3” (75 mm) at each side for proper venting.
3. Adequate support for the weight; 55 lb. (25 kg.).
4. Access at the rear for making the pneumatic and electrical connections.
5. Access at the front for occasional scrubber replacement.
6. Operate the M701H with the cover on.