5.6 Replacing the NO - NO2 Scrubber Media
This procedure is identical to the charcoal scrubber replacement procedure except
that the canister should be refilled with Purafil .
NOTE
After replacing the material in these scrubbers it can take up to 48 hours for
the material to dry out before the 751H is functioning optimally.
5.7 Replacing the CO Scrubber
The CO scrubber is attached to the bulkhead adjacent to the inst rument panel
(Figure 4-2). This is not a heated scrubber. It is secured with f our screws through the
body of the scrubber into the panel.
1. Turn off power and unplug the 751H.
2. Remove the Charcoal and NO-NO2 scrubbers to gain access to the hardware.
3. Disconnect the tubing and remove the four screws to remove the scrubber.
CAUTION
THE SCRUBBER MAY BE HOT.
4. With a wrench, remove the two fittings from the top of the scrubber cartridge.
5. Pick out the retaining screens.
6. Shake out the catalyst beads and dispose. No special disposal metho ds
required.
7. Pour in new catalyst to 1/2” from the top of the bores. Tap the cartridge sides
gently to settle the beads and top up to the 1/2” level.
8. Return the retainer screens.
9. Replace the TFE tape on the two unions and replace the union s in the cartridge.
10. Reassemble the scrubber and reattach it to the panel.
11. Reconnect the tubing and receptacle.
12. Turn on the 751H and perform a leak check (see Section 5.11).