The spare cell should be carefully stored in an area that is not subject to large variations in ambient temperature (75 Deg. F nominal), and in such a way as to obviate any possibility of damage. Under no circumstances, disturb the integrity of the cell package until the cell is to be actually used.

No tools are required to replace the cell in the instrument. Simply unscrew (ccw) the plug at the bottom of the analyzer and the cell will drop out of the manifold cavity.

Remove the new cell from its package, and carefully remove the shorting device. Place the cell on the end of the manifold plug -- so that the sensing surface of the cell is in contact with the plug and the electrical contact plate end of the cell is facing upwards. Insert the cell and plug in the manifold cavity, and screw the plug back into place. Apply as much pressure as you can with your fingers, but use no tools.

After the cell has been installed, purge the instrument with an oxygen free gas (or the sample), and then proceed as directed in Section 3.1 and 3.2.1.

4.4Cell Warranty. The Class B-2CXL cell employed in the Model 311XL is warranted for six (6) months of service.

With regard to spare cells, service time starts when the cell is removed from its shipping package. The customer should stock only one spare cell per instrument at a time. Do not attempt to stock- pile spare cells.

The Model 311XL should not be used in applications where CO2 is a major component in the sample. Concentrations of 1,000 ppm or less will not effect the cell performance. The following page is a graph showing the effects of CO2 on cell life.

If a cell was working satisfactorily, but ceases to function before the warranty period expires, the customer will receive credit, toward the purchase of a new cell.

Customer’s having warranty claims must return the cell in question to the factory for evaluation, after obtained an RMA number. If it is determined that failure is due to faulty workmanship or material, the cell will be replaced at no cost to the customer.

-9-

WARNING: Evidence of damage due to tampering or mishandling will render the cell warranty null and void.

Page 13
Image 13
Teledyne ECO#02-0227 instruction manual

ECO#02-0227 specifications

Teledyne ECO#02-0227 represents an advancement in environmental monitoring technologies, specifically designed to address the ever-evolving requirements of water quality analysis. This innovative device is equipped with cutting-edge features that enhance its functionality, providing users with reliable data for a variety of applications ranging from academic research to industrial processes.

One of the standout features of the Teledyne ECO#02-0227 is its ability to measure multiple parameters simultaneously. It offers real-time monitoring of critical water quality indicators, including dissolved oxygen, turbidity, pH levels, and chlorophyll concentrations. The simultaneous data collection capability accelerates decision-making processes, allowing for timely interventions in environmental management.

The device leverages advanced sensor technology, ensuring high accuracy and precision in measurements. The sensors are designed to minimize interference and provide consistent data across a range of conditions. This is particularly important for applications where water quality can fluctuate significantly, such as in rivers, lakes, and wastewater treatment facilities.

Another notable characteristic of the ECO#02-0227 is its rugged design, which makes it suitable for deployment in harsh environments. It features waterproof housing and is resistant to corrosion, ensuring durability and longevity even under extreme conditions. This robustness makes the device an ideal choice for field applications, where reliability is paramount.

Moreover, the Teledyne ECO#02-0227 utilizes smart connectivity options, enabling seamless data transfer and remote monitoring capabilities. Users can easily access real-time data via wireless networks, integrating the data into broader environmental monitoring systems. This connectivity facilitates collaboration among researchers and regulators, promoting more effective management of water resources.

In addition to its technical specifications, the device is user-friendly, with a straightforward interface that allows for easy navigation through settings and data retrieval. Comprehensive documentation and support resources ensure that users can maximize the device's potential without extensive training.

Overall, the Teledyne ECO#02-0227 stands out in the field of water quality monitoring. Its multi-parameter capabilities, advanced sensor technology, robust design, smart connectivity, and user-friendly interface make it an invaluable tool for environmental professionals seeking reliable and accurate data to support their initiatives. With this device, users can monitor and protect water quality more effectively, contributing to healthier ecosystems and communities.