OverviewTeledyne API – Model T300/T300M CO AnalyzerTeledyne API – Model T300/T300M CO Analyzer

Teledyne Analytical Instruments 88
O2 Range 1 O2 RANGE % The range setting for the optional O2 Sensor.
CO2 Range2 CO2 RANGE % The range setting for the optional CO2 Sensor.
CO Measure CO MEAS MV The demodulated, peak IR detector output during the measure
portion of the GFC Wheel cycle.
CO Reference CO REF MV The demodulated, peak IR detector output during the reference
portion of the GFC Wheel cycle.
Measurement /
Reference Ratio MR Ratio -
The result of CO MEAS divided by CO REF. This ratio is the
primary value used to compute CO concentration. The value
displayed is not linearized.
Sample Pressure PRES In-Hg-A The absolute pressure of the Sample gas as measured by a
pressure sensor located inside the sample chamber.
Sample Flow SAMPLE FL cm3/min Sample mass flow rate as measured by the flow rate sensor in the
sample gas stream.
Sample
Temperature SAMP TEMP C The temperature of the gas inside the sample chamber.
Bench
Temperature BENCH TEMP C Optical bench temperature.
Wheel
Temperature WHEEL TEMP C GFC Wheel temperature.
Box Temperature BOX TEMP C The temperature inside the analyzer chassis.
O2 Cell
Temperature1 O2 CELL TEMP3 C The current temperature of the O2 sensor measurement cell.
Photo-detector
Temp. Control
Voltage
PHT DRIVE mV The drive voltage being supplied to the thermoelectric coolers of the
IR photo-detector by the sync/demod Board.
Slope SLOPE - The sensitivity of the instrument as calculated during the last
calibration activity.
Offset OFFSET - The overall offset of the instrument as calculated during the last
calibration activity.
O2 Sensor
Slope 1 O2 SLOPE - O2 slope, computed during zero/span calibration.
O2 Sensor Offset
1 O2 OFFSET - O2 offset, computed during zero/span calibration.
CO2 Sensor
Slope2 CO2 SLOPE - CO2 slope, computed during zero/span calibration.
CO2 Sensor
Offset 2 CO2 OFFSET - CO2 offset, computed during zero/span calibration.
Current Time TIME - The current time. This is used to create a time stamp on DAS
readings, and by the AUTOCAL feature to trigger calibration events.
1 Only appears when the optional O2 sensor is installed.
2 Only appears when the optional CO2 sensor is installed.
3 Only available on the GFC7001T.
4.1.2. WARNING MESSAGES

The most common instrument failures will be reported as a warning on the analyzer’s

front panel and through the COMM ports. Section 12.1.1 explains how to use these

messages to troubleshoot problems. Section 4.1.2 shows how to view and clear warning

messages.

Table 4-3: List of Warning Messages

MESSAGE MEANING
ANALOG CAL WARNING The instrument’s A/D circuitry or one of its analog outputs is not calibrated.