8. Device Status

8.1.ION GAUGE DEVICE STATUS: 4-DIGIT OCTAL WORD

When requested to transmit its status the IGE-3000 responds with a 4-digit octal word.

FIGURE 7 -1 – DIGIT OCTAL WORD

bit 11: Serial Receiver Overload

bit 10: Main Board EEPROM Error.

bit 9: Communication Syntax Error.

bit 7 Emission Setting.

bit 8 Emission Setting.

bit 6: Filament Selection. (0=filament 1, 1=filament 2)

FIGURE 8- BITS 7 & 8 EMISSION CODE

BITS 7 & 8

BIT 8

BIT 7

NO EMISSIONS

0

0

0.01 Ma

0

1

0.1 mA

1

0

1.0 mA

1

1

bit 5: Filament Status. (0=non-operating, 1=operating)

bit 4: Degas (0=degas off, 1=degas on)

bit 3: Pressure less than (0.1e-9 Torr) Set if true.

bit 2: Pressure Greater than (1.0e-3 Torr) set if true.

bit 1: High Set point Alarm Set if pressure is greater than the high set point

bit 0: Low Set point Alarm Set if pressure is less than the low set point

8.2.Command Syntax

In the following examples of syntax codes, the special characters are explained:

The characters in square brackets [ ] represents a command string, either upper or lower case command characters accepted. All characters must follow each other in the string with no spaces or other characters.

The characters within wavy brackets { } contain choices for the appropriate command.

The characters within the symbols < > are the common abbreviations for the one digit ASCII control codes which they represent, (e.g. <CR> represents carriage return).

169-082010_Ion-Gauge-3000

Page 18 of 21

Page 18
Image 18
Teledyne IGE 3000 instruction manual ION Gauge Device Status 4-DIGIT Octal Word, Command Syntax