2. General Information

2.1.GENERAL INFORMATION

The Teledyne Hastings Instruments IGE-3000 is a small, fully self contained ion gauge electronics module. The mini- Bayard Alpert style ion gauge tube is attached to the electronics module by a 12 pin circular connector. The IGE-3000 is designed for quick, easy installation and will provide the user with long lasting, trouble free, reliable vacuum measurement.

2.2.IGE-3000 Mini Bayard/Alpert (B/A) Tube

The IGE-3000 mini B/A tube is all metal construction, with the primary enclosure fabricated of 304

stainless steel. It is UHV compatible and utilizes a fine wire collector located in the center of the grid. The duel Yttria coated filaments are individually supported to provide uniform output over variations in temperature and normal ageing. The typical user should experience a filament lifetime of approximately 10,000 hours. Its physical size is much smaller than the traditional glass type B/A tube. Since there is no glass envelope, the traditional safety concerns associated with glass type B/A tubes is eliminated.

2.3. Specifications.

Pressure Range

1x10-2 to 1x10-10 Torr

*Accuracy

±15% of reading for the range 1x10-2 to 1x10-8 Torr

Degas

5W e-beam

Emission Current

0.01, 0.1, 1 mA, or Auto

Filament Bias Potential

+30 V ±0.3%

Grid Potential

+180 V ±0.3%

Operating Temperature

0-50°C

Digital Communications

RS232/485

Connector

15-pin subminiature D

Bake out Temperature

200°C with electronics removed

** Electrical Noise Immunity

ENC 61010 Class A and IEC 61326-1: Section 6.2

Input Voltage Operating Range

11.5 VDC to 30 VDC

Input Power

24 VDC @ 350 mA (Normal Operation)

.............................................................................................................

 

 

 

 

 

 

24 VDC @ 850 mA (During Degas)

Power/Interface Connector

15 pin subminiature

Process Control

2 independent TTL set point outputs

Weight

1.5 lbs. (max)

Transducer Mounting

Any position

Remote Control Capable

Automatic High Pressure Cut off

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

*Accuracy specification is with the B/A tube originally shipped with the IGE 3000. The IGE 3000 is calibrated with and matched to the B/A tube at the factory. Changing or swapping the B/A tube with a B/A tube one other than the one originally shipped with the unit may effect the calibration.

**Complies with IEC 61010-1 (Safety requirements for measurement, control, and laboratory use).

169-082010_Ion-Gauge-3000

Page 6 of 21

Page 6
Image 6
Teledyne IGE 3000 instruction manual General Information, IGE-3000 Mini Bayard/Alpert B/A Tube, Specifications