Service Manual MX300 Teledyne Analytical Instruments 40

Symptom Why What To Do

The sensor does not
react to changes in
oxygen concentration,
or the readings are
unstable and drifting.
Water is condensing
on the sensing
surface.
Electrical interference
is disrupting the
electronics
A) Remove the sensor from tee
adapter and unscrew the
plastic flow diverter. Using
absorbent tissue or cotton
swab, gently wipe off sensing
surface inside threaded
portion of sensor assembly.
B) Relocate unit away from
sources of electrical noise
such as cauterizing equipment
and two-way radios.
The display is flashing

SENSOR

The unit has detected
a fault in the signal
from the sensor.
Sensor has expired.
The sensor has been
exposed to a gas
containing little or no
oxygen.
A) Check sensor cable
connections and make sure
they are completely inserted
into the mating connector
and the capture nut is firmly
in place.
B) Expose the sensor in 100%
O2 and check calibration.
C) The sensor output has fallen
to a level where it is no longer
usable. Replace sensor.
The oxygen reading
fluctuates or appears to
be incorrect.
Like all O2 sensors,
the R17MED detects
the changes in the
partial pressure of
O2.
A) During calibration, make
sure there is no restrictions
on exhaust side of sensor. If
the reading changes with
flow, the sensor is
pressurized or there may be
a leak in the system.
B) If a high degree of accuracy
is desired, or the
concentration of O2 is in
excess of 40%, calibration
with 100% is recommended.
C) If humidified gas is used to
ventilate the patient, water
vapor actually dilutes the
gas. See Appendix:
Humidity, Temperature.
D) If a blender is used, check
its calibration. See
Appendix: Discrepancy in
Readings.