Recalibrate after replacing the sensor.
Use properly installed alkaline batteries only.
Make sure the R17MED sensor is properly attached.
Remove the plastic flow diverter only when using the tee
adapter.
Remove and save the plastic flow diverter when using the
sensor in non-flowing applications (incubators, tents,
etc.)
Perform an alarm test before each use (press the ALARM
TEST key).
Clean the case with isopropyl alcohol or mild detergent
only.
– DON’T –
Use this monitor if you suspect any malfunction.
Use the instrument in the presence of flammable gases.
Use anything but alkaline batteries.
Autoclave or freeze the sensor or instrument.
Open or try to repair a leaking or broken sensor.
Immerse the unit or sensor in any liquid.
Pass hot or cold gas mixtures over the sensor.
Adjust the reading in air after 100% calibration
Expose the unit to devices that produce high levels of
radio, short wave, microwave, x-ray, or high frequency
interference.
Use cleaning agents or liquids in the cable receptacles or
around the battery compartment.
Place the unit itself in a water vapor-saturated
environment.