Ultra Trace Oxygen Analyzer Operation 4
4-19
Teledyne Analytical Instruments
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in section
4.3.3 to enter your password. Once you have clearance to proceed, enter the
Alarm
function.
Press the
Alarm
button on the front panel to enter the
Alarm
function.
Make sure that AL1 is blinking.
AL1 AL2
Choose Alarm
Set up alarm 1 by moving the blinking over to AL1 using the < >
arrow keys. Then press Enter to move to the next screen.
AL1 10.000 ppm HI
DftN FsN LtchN
Five parameters can be changed on this screen:
Value of the alarm setpoint, AL1 #### ppm (oxygen);
value can be set from 0 to 999 ppb + 1.000-1000.00 ppm.
Out-of-range direction, HI or LO
Defeated? DftY/N (Yes/No)
Failsafe? FsY/N (Yes/No)
Latching? LtchY/N (Yes/No).
To define the setpoint, use the < > arrow keys to move the
blinking over to AL1 ####. Then use the ∆∇ arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, the setpoint units are
ppm O2.)
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the ∆∇ arrow
keys to change the parameter.
Once the parameters for alarm 1 have been set, press

Alarms

again, and repeat this procedure for alarm 2 (AL2).
To reset a latched alarm, go to Dft and then press either two
times or two times. (Toggle it to Y and then back to N.)
OR
Go to Ltch and then press either two times or two times.
(Toggle it to N and back to Y.