3 Installation Model Ultra Trace 3000
3-2 Teledyne Analytical Instruments
Figure 3-1: Front Panel of the Model Ultra Trace 3000
All operator controls are mounted on the control panel, which is hinged
on the left edge and doubles as the door that provides access to the sensor
and cell block inside the instrument. The door is spring loaded and will
swing open when the button in the center of the latch (upper right corner) is
pressed all the way in with a narrow gauge tool (less than 0.18 inch wide),
such as a small hex wrench or screwdriver Allow clearance for the door to
open in a 90-degree arc of radius 7.125 inches. See Figure 3-2.
Provide 1 inch of air clearance at the bottom and sides of the instrument
for proper air flow for TEC fun.
Failure to provide such a clearance will cause damage to the TEC
system. Do not restrict this openings.
Figure 3-2: Required Front Door Clearance
2.0
1.0
Hinge
Latch
Analyze System
Span Zero
Alarms Range
Standby
Enter Escape Set Flow
ULTRA TRACE 3000
ULTRA TRACE ANALYZER
6.7”
10”