A-2
Appendix Model Ultra Trace 3000
Teledyne Analytical Instruments
Power: Universal power supply 85-250 V ac, at
47-63 Hz.
Operating Temperature: 5-40 °C
Accuracy: ±2% of full scale for all ranges except 0-250
ppb range, at constant temperature.
For 0-250 ppb range accuracy is as follows:
±5 ppb at constant temperature.
±10 ppb over temperature variation of 20° -
30°C.
±15 ppb over temperature variation of 15° -
35°C.
All accuracy specifications are contingent
upon the completion of zero ans span calibra-
tion.
All accuracy is established at constant pres-
sure and equilibrium has been established.
Analog outputs: 0-1 V dc percent-of-range,
0-1 V dc range ID.
4-20 mA dc (isolated) percent-of-range,
4-20 mA dc (isolated) range ID.
Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep (6.96
in high, 8.7 in wide, 12.2 in deep).