A-1

Ultra Trace Oxygen Analyzer Appendix
Teledyne Analytical Instruments
Appendix

A-1 Specifications

Packaging: General Purpose
Flush panel mount (Standard).
Relay rack mount. Contains one
instrumentsin one 19" relay rack mountable
plate (Optional).
Sensor: Teledyne B-2CXL trace analysis Micro-Fuel
Cell.
Cell Block: 316 stainless steel.
Sample System: All wetted parts of 316 stainless steel.
90 % Response Time: Less than 30 seconds at 25 °C (77 °F) on 10
ppm range.
120 seconds on 250 ppb range, programmable
response in ppb range.
Ranges: Three user definable ranges from 0250 ppb to
01,000 ppm, plus over range of 0-1,000 ppm.
Autoranging with range ID output.
Alarms: One system-failure alarm contact to detect
power failure or sensor-zero and span failure.
Two adjustable concentration threshold alarm
contacts with fully programmable setpoints.
Displays: 2-line by 20-character, VFD screen, and one 5
digit LED display.
Digital Interface: Full duplex RS-232 communications port.