Sure-Flow Electrode Cleaning
Cat. No. 9678BNWP and 967800
If the area between the electrode sleeve and inner cone becomes clogged with sample or precipitate, flush the area with filling solution or distilled water.
1.Hold the electrode body with one hand and use your thumb to push down on the electrode cap to drain the chamber. Push down on the cap until all the filling solution is drained from the chamber.
2.Fill the electrode with distilled water and then push down on the cap until all the water is drained from the chamber.
3.Fill the electrode with fresh filling solution up to the fill hole. Push down on the cap to allow a few drops of filling solution to drain out of the electrode and then refill any lost filling solution.
Disassembling the Sure-Flow Electrode
Note: Disassembly is usually not required and should not be done unless a thorough cleaning is required.
1.Tip the electrode so the filling solution moistens the
2.Unscrew the cap counter
3.Hold the outer sleeve with one hand and firmly push down on the threaded portion with the thumb and forefinger to separate the inner body from the sleeve.
4.Grasp the cone with a clean,
16 | Redox/ORP Electrode User Guide |