17Redox/ORP Electrode User Guide

Reassembling the Sure-Flow Electrode
1. Moisten the O-ring on the electrode body with a drop of
filling solution. Insert the screw-thread end of the electrode
body into the tapered, ground end of the sleeve.
2. Push the body into the sleeve using a gentle twisting
motion until the bottom surface of the inner cone is flush
with the tapered end of the sleeve.
3. Place the spring onto the electrode body and screw on the
cap. Refill the electrode with filling solution.
Electrode Storage

Short-term Storage (up to one day)

To ensure a quick response and free-flowing junction, the
reference junction must not dry out. It is recommended that
the electrode be soaked in a 4 M potassium chloride (KCl)
solution. Electrodes may also be stored in water.

Long-term Storage (over one day)

For gel-filled electrodes– rinse the electrode with distilled
water, dry the electrode with a lint-free tissue and cover the
measuring surface with the protective cap.
For refillable electrodes– rinse the electrode with distilled
water, drain all of the filling solution from the electrode, flush
out the electrode with distilled water, cover the measuring
surface with the protective cap and store the electrode dry.