Topcom BPM 1900f

3 Technical specifications

 

 

ENGLISH

Power source

6V DC: 4 x LR06 (AA) batteries

 

Measurement method

Oscillometric

 

Measurement zone

Pressure: 20~280 mmHg

 

 

Pulse: 40~180 beats/minute

 

Accuracy

Pressure: ±3 mmHg

 

 

Pulse: within ±5% of reading

 

Pressure sensor

Semi-conductor

 

Inflation

Pump driven

 

Deflation

Automatic Pressure release valve

 

Memory capacity

90 memories

 

Auto-shut-off

60 seconds after last key operation

 

Operation environment

Temperature : 5°C~45°C(41°F~113°F)

 

 

Humidity : 40~85% RH max.

 

Storage environment

Temperature : -10°C~60°C (14°F~140°F)

 

 

Humidity : 10~95% RH max.

 

Weight

400g (without batteries)

 

Arm circumference

24~36 cm

 

Limited patient

Over 18 years old

 

Dimensions

150(L) x 100(W) x 68(H) mm

 

K.B.

Short time operation 2 minutes.

 

Classification

Classification of type BF

 

 

 

 

*Specifications are subject to change without notice.

4 Helpfull tips

Here are a few helpful tips to help you obtain more accurate readings:

Do not measure your blood pressure immediately after consuming a large meal. To obtain more accurate readings, please wait one hour before measuring.

Do not smoke or drink alcohol before measuring your blood pressure.

You should not do be physically tired or exhausted while taking a measurement.

It is important that you relax during measurement. Try to take 15-minutes rest before a reading.

Do not take measurements if you are under stress or under tension.

Take your blood pressure at normal body temperature. If you are feeling cold or hot, wait a while before taking a measurement.

If the monitor is stored at very low temperature (near freezing), have it placed at a warm location for at least one hour before using it.

Wait about 5 minutes before taking the next pressure measurement.

5

Page 5
Image 5
Topcom 1900f manual Technical specifications, Helpfull tips