Tripp Lite HCRK-1 Connection / Start-Up continued, Turn the Power Supply Module’s Outlets ON

Models: HCRK-1

1 12
Download 12 pages 23.6 Kb
Page 6
Image 6
Connection / Start-Up continued

Connection / Start-Up continued

5Turn the Power Supply Module’s Outlets ON

Press and hold the “Power” button* for one second. The alarm will beep once briefly after one second has passed. Release the button. The “Power” LED* will illuminate to indicate that the Power Supply Module’s outlets are ON. The Battery Charge LED Meter* will indicate the approximate charge level of the connected Battery Module.

* Located on the Remote User Interface (RUI).

5

6 Plug Your AC Equipment into

 

the Power Supply Module’s

 

Outlets

 

The Power Supply Module provides filtered,

 

pure sine-wave 120V AC power for a variety of

 

equipment—computers, monitors, diagnostic

 

devices, scanners and more. You will overload

 

the Power Supply Module if the total VA ratings

 

for all the equipment you connect exceeds

6

200VA (the Power Supply Module’s continuous

 

output capacity).*

 

*To find your equipment’s VA ratings, look on their nameplates. If the equipment is listed in amps, multiply the number of amps by 120 to determine VA. (Example: 1 amp x 120 = 120 VA).

7

OPTIONAL—Install PowerAlert

 

 

Data-Saving Software

 

 

This connection is optional. The Power

 

 

Supply Module and Battery Module will

 

 

function properly without this connection.

 

 

Using the included cable*, connect the

 

 

Power Supply Module’s USB port to the

 

 

USB port on any laptop or desktop

 

 

computer mounted on the mobile cart.

7

 

Insert the included PowerAlert CD and load

 

 

the version of PowerAlert appropriate for

 

 

the computer’s operating system. Follow

 

 

the setup and configuration screens for your

 

 

specific application.

 

*If desired, you can substitute standard USB cable in place of the included cable.

6

Page 6
Image 6
Tripp Lite HCRK-1 Connection / Start-Up continued, Turn the Power Supply Module’s Outlets ON, Plug Your AC Equipment into