8. Press the Down Arrow to advance the cursor to the Sta Dly: field. Use the Left and Right Arrow Keys
to navigate between the Hour and Minute fields. Use Input Dial to set the Station Delay duration. Station delay is the wait period when a station finished watering and when the next station is activated. In most situations, station delay is used to allow slow
9. | Press the Down Arrow | to advance the cursor to the Syringe: field. Use the Input Dial |
|
| to set | ||||||
| |||||||||||
| the Syringe duration. When a syringe duration | ||||||||||
| time number to indicate that this start time is a syringe start. During a syringe, the stations will all activate for | ||||||||||
| the syringe duration regardless of their assigned runtimes. For this example, do not set a Syringe duration. | ||||||||||
10. | Press the Down Arrow | to advance the cursor to the Repeats#: field. Use the Input Dial |
|
| to | ||||||
| |||||||||||
| enter the number of times the program will repeat the cycle. For this example,set Repeat to remain at 00. | ||||||||||
11. | Press the Down Arrow | to advance the cursor to the Soak#: field. Use the Input Dial |
|
|
|
|
|
| to enter | ||
| |||||||||||
| the soak duration before the station is reactivated. For this example, set a soak duration of 00:10:00. |
| |||||||||
12. | Press the Down Arrow | to advance the cursor to the Sta#: field. Use the Input Dial |
|
|
|
|
|
|
|
| to select |
| |||||||||||
| the first station being irrigated in the sequence. For this example, select station 06. |
|
13.Press the Right Arrow to advance the cursor to the next value. This value will indicate the last station in the sequence. If irrigating only one station, this value will be the same as the first value. For this example, select station 10. With 6 being the first station and 10 being the last station, the program schedule will water stations 6, 7, 8, 9 and 10 in sequentially.
14.Press the Right Arrow to advance the cursor to the next entry field. This entry field will indicate the
runtime in hours, minutes and seconds ( HH:MM:SS). Use the Input Dial and the Right Arrow Key to select the appropriate runtime value. For this example, set the value to 00:15.
Repeat Steps
15. Press the Down Arrow | to advance the cursor to the Simult: field. Use the Input Dial |
| to select |
| |||
the maximum allowable simultaneous active station. For this example, set the value to 05. |
|
|
Note: The Program’s Simultaneous setting is limited to the System’s Simultaneous value.
When finished, the display should read:
P02 100% 01:15:00
Hold: None
Type: Standard
Days: SMTWTFSSMTWTFS set>:
Start : 01 05:45am
Sta Dly: 00:00 (h:m)
Syringe: 00 min
Repeats: 00
Soak: 00:10:00
Sta#:
Sta#:
Sta#:
Simult: 05
12