Toro TDC-100/200 manual P02 100% Hold None Type Standard, When finished, the display should read

Page 12

8. Press the Down Arrow to advance the cursor to the Sta Dly: field. Use the Left and Right Arrow Keys

to navigate between the Hour and Minute fields. Use Input Dial to set the Station Delay duration. Station delay is the wait period when a station finished watering and when the next station is activated. In most situations, station delay is used to allow slow -shutting valves time to fully close before activating another station and also to allow the water source, such as a well, to recover. For this example, set Station Delay to remain at 00.

9.

Press the Down Arrow

to advance the cursor to the Syringe: field. Use the Input Dial

 

 

to set

 

 

the Syringe duration. When a syringe duration (0–99 minutes) is set, an asterisk is placed in front of the start

 

time number to indicate that this start time is a syringe start. During a syringe, the stations will all activate for

 

the syringe duration regardless of their assigned runtimes. For this example, do not set a Syringe duration.

10.

Press the Down Arrow

to advance the cursor to the Repeats#: field. Use the Input Dial

 

 

to

 

 

enter the number of times the program will repeat the cycle. For this example,set Repeat to remain at 00.

11.

Press the Down Arrow

to advance the cursor to the Soak#: field. Use the Input Dial

 

 

 

 

 

 

to enter

 

 

the soak duration before the station is reactivated. For this example, set a soak duration of 00:10:00.

 

12.

Press the Down Arrow

to advance the cursor to the Sta#: field. Use the Input Dial

 

 

 

 

 

 

 

 

to select

 

 

the first station being irrigated in the sequence. For this example, select station 06.

 

13.Press the Right Arrow to advance the cursor to the next value. This value will indicate the last station in the sequence. If irrigating only one station, this value will be the same as the first value. For this example, select station 10. With 6 being the first station and 10 being the last station, the program schedule will water stations 6, 7, 8, 9 and 10 in sequentially.

14.Press the Right Arrow to advance the cursor to the next entry field. This entry field will indicate the

runtime in hours, minutes and seconds ( HH:MM:SS). Use the Input Dial and the Right Arrow Key to select the appropriate runtime value. For this example, set the value to 00:15.

Repeat Steps 8–10 to assign stations 10–20 with a runtime of 10 minutes.

15. Press the Down Arrow

to advance the cursor to the Simult: field. Use the Input Dial

 

to select

 

the maximum allowable simultaneous active station. For this example, set the value to 05.

 

 

Note: The Program’s Simultaneous setting is limited to the System’s Simultaneous value.

When finished, the display should read:

P02 100% 01:15:00

Hold: None

Type: Standard

Days: SMTWTFSSMTWTFS set>:

Start : 01 05:45am

Sta Dly: 00:00 (h:m)

Syringe: 00 min

Repeats: 00

Soak: 00:10:00

Sta#: 06-10 00:15

Sta#: 15-20 00:10-

Sta#: ––––––– ––:––

Simult: 05

12

Image 12
Contents TDC-100/200 User’s Guide Page Index Introduction General EditingTiming Mechanism TDC-100/200Mother Board Mother Board ComponentsDaug hter Board Daughter Board ComponentsPower-Up Diagnostics TDC-100/200 System BootingHome Key Rev D1.21, 04/25/06 Detect200sta, 00sen PD 05/24/06System Settings UnitsDisplay Adj Sensor Reset All Programs To defaults Stop to escape Reset All Stations To defaults Stop to escape Reset All Defaults Stop to escape Reset Sta Disables Stop to escapeStation Settings Standard Irrigation Program Setting Scheduled WateringPress the Down Arrow To selectWhen finished, the display should read P02 100% Hold None Type StandardGrow In Irrigation Program Setting 0400pmManual Watering Manual M-Manual Sta# 01-12 Simult Press Start to waterMulti-Manual Station Activation Directions Manual Syringe Runtime 02 min Program Press Start to water Manual Start Prog Program Press Start to waterManual Syringe Activation Directions Manual Program Activation DirectionsPercent Adjust Percent Adjust Global100% Program P01 100% Station S01 100%Multi-Manual Starting Stop to cancel Syringe Starting Program Stop to cancelPause / Resume Key Resume Prg Press Pause to pause 01 Programs PausedPause / Resume Function Directions Cancelling P01 Stop to escape Stop KeyStop Function Directions Diagnostics Fuse & Circuit Breaker Specifications Controller SpecificationsElectromagnetic Compatibility