7.4 Arbitrary Waveform Mode

 

 

 

1

 

4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Arb Mode Int

 

Clock

 

 

Phase: 126.35 deg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

 

 

Cont Clock: 1,000,000.00 Hz

10,000 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 7.4-1: Arbitrary Waveform Mode display

1. Int Clock / Ext Clock / Lock Master Lock slave mode

2

5

In this field select the clocking mode for the arbitrary waveform system:

Internal Clock - The sample clock is generated internally. The clock frequency is entered directly via the front panel.

External Clock - The sample clock (40 MHz max) is taken from the External Arb Clock input on the back panel.

*** Warning ***

TTL/CMOS levels (0V to +5V) only.

Lock Master - The unit provides the master timing signals for locking multiple 4070A Arb generators together.

Lock Slave - The unit receives all timing signals from a master unit. Used when locking multiple 4070A Arb’s together. For more information on locking Arb’s together, see the section 7.9.

2. Phase

This field is used during Multiple Unit Locking and only appears when the unit is in Lock Slave mode. It allows you to specify a phase offset of the arbitrary waveform relative to the waveform being generated by the Master Unit. For more information, see section 7.9.

3. Continuous/Triggered mode

In this field select whether the waveform is generated continuously or on a single event basis. If you select Continuous mode, then the waveform is immediately restarted once the last point is reached. If you select a Triggered mode, then the waveform generation is halted after the last point is reached, and the waveform is not restarted again until another trigger occurs. The trigger can come from three sources:

1.Pressing the Trigger key

2.Applying a low-to-high transition on the Ext. Trig In connector

3.Sending an ASCII "T" to the RS-232 port

The 4070A will simultaneously accept a trigger from all of the above sources. To set Continuous-triggering mode, press 1.

To set Triggered mode, press 0.

Pressing any arrow key or rotating the wheel will toggle the run mode between Continuous and Triggered.

4. Clock Frequency

In this field enter the Arbitrary Waveform clock frequency. This is the frequency at which your waveform values are sent to the Digital to Analog converter (DAC). You may enter from 0 Hz (DC) to 40.0 MHz in 0.01 Hz steps.

If a value of 0 Hz is entered, the arbitrary waveform clock is halted.

5. Level

In this field enter the output level, from 4 mVp -p to 10 Vp -p in 1 mV steps. This level is the peak-to-peak voltage swing of the waveform across a 50O loaded, into an open circuit, the voltage swing will be twice the specified peak- to-peak value.

Offset

Enter a DC offset voltage by pressing the Offset key. For more information refer to section 4.4.

49

BK Precision 4070A User Manual Rev.2.2

Image 54
Contents User’s Manual Page Model 4070A User’s Manual +K Precision CorpTable of Contents Arbitrary waveform system Warranty informationRemote operation DC Operation OptionIntroduction Description Feature summary ModesInstallation and setup Sync out connectorSIG Out connector External Modulation In connector External Trigger/Gating/FSK/BPSK input RS232 Interface connector External Reference Input / Ext Arb. Clock connectorExternal Arb Sync In / Out connector Line in connectorField arrow keys Mode keyOperating controls and keys Recall / Store keyTrigger key Numeric keys 0 to 9,É, ê, è, ç keys One Touch Mode Keys Clear keyMHz/dBm, KHz/Vp-p/Sec, Hz/mVp-p/mS keys Changing frequency Operating mode selectionOperating guide Changing levelChanging values Modifying an existing valueEntering a new value Mode descriptions Basic Sinewave CW ModeSinewave Mode 000,000.00 Hz 10.0 dBm Internal AM Mode External AM Input gain 000,000.00 Hz PEP level -10.0 dBm External AM ModeExternal AM Mode Parameters Input GainInt FM 1,000 Hz Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm Internal FM ModeInternal FM Mode Parameters Peak Frequency DeviationExternal FM Mode External FM Mode ParametersExternal FM Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm Peak Phase Deviation Internal PM ModeInternal PM Mode Parameters Int PM 1,000 Hz Pk dev 180 deg 000,000.00 Hz 10.0 dBmExternal PM Mode External PM Mode ParametersExternal PM Pk dev 180 deg 000,000.00 Hz 10.0 dBm Sweep Mode Up / Down sweep Sweep timeMark Frequency Internal FSK ModeInternal FSK Mode Parameters Space FrequencyExt FSK Mark 1,000,000.00 Hz Space 2,000,000.00 Hz 10.0 dBm External FSK ModeExternal FSK Mode Parameters Ext FSK InputBurst Mode Introduction Burst Mode Trig Ton 10,000 mS 000,000.00 Hz Off 10.0 dBmBurst frequency Int SSB 1,000 Hz Upper Sideband 000,000.00 Hz 10.0 dBm Internal SSB ModeInternal SSB Mode Parameters Upper/Lower Sideband selectionsExternal SSB Mode External SSB Mode ParametersExternal SSB Upper Sideband 000,000.00 Hz +10.0 dBm Dtmf Generation Mode Parameters Dtmf Generation ModeDtmf generation # Duration 200 mS Delay Level -10.0 dBm BackgroundDtmf Delay Dtmf DurationDtmf detection on Ext Mod input 3 # a Dtmf Detection ModeDtmf Detection Mode Parameters Dtmf detected digitsPower Level Power & Voltage Measurement ModePower & Voltage Measurement Mode Parameters Voltage LevelArbitrary Mode Remote ModeOther mode Internal Bpsk Mode Internal Bpsk Mode ParametersInt Bpsk 1,000 Hz 000,000.00 Hz 10.0 dBm External Bpsk Mode External Bpsk Mode ParametersExternal Bpsk 000,000.00 Hz 10.0 dBm Dualtone Generation Mode 10,000 Hz F2 Ton 000 mS Toff 10.0Dualtone Examples Data Modulation Mode Parameters Data Modulation ModeUsing the Data Modulation Mode Baud RateDownloading Modulation Message Data Example MessageTriggering the Transmission Saving the MessageStart End 1,000,000.00 Hz 10.0 dBm Voltage Controlled Oscillator VCO ModeVCO Mode Parameters End FrequencyConnecting Remote OperationChecking your connection with Hyper Terminal Remote Control Operation Remote Control Programming Rules× Ø Remote Control CommandsReset 4070A to Sine wave mode Returns CK1.0 Enable, Disable front panel keys and rotary knob E1,0 Enable, Disable LCD echo to terminalReport hardware and software versions ? or H Print help menu Linear Trig Up Time 10,000 mSF0-9 Move cursor to field 0 to Return a CRemote Control Examples ExampleArbitrary Waveform Mode Quick Start GuidePin 9-pin cable Pin 25 pin cable adapter Introduction to the Arbitrary Waveform System Description of the Arbitrary Waveform GeneratorFunction Generator Switching ModesFeature Summary Pulse GeneratorPhase Arbitrary Waveform ModeInt Clock / Ext Clock / Lock Master Lock slave mode Continuous/Triggered modeWaveform Function Generator ModeFunction Gen Wave Pos Ramp Rep Freq 100,000 Hz 000 mV Repetition FrequencyPulse Generator Mode Positive OnlyDuty Cycle Saving Arbitrary Waveforms Using WAVELOAD.EXEDownloading arbitrary waveforms Cwaveload /COMx /y /baudz filename.extUsing your own program Data Formats Floating Point FormatRules for Floating Point format Time & Value Floating Point Format Example #2Digital Format Integer Format Hexadecimal Format Rules for Hexadecimal formatExample The data value E468 is handled as follows Binary Format Rules for Binary formatExample Multiple Units Locking ArbGenerator Starting Address = Number of waveform points * Phase Offset More on Phase OffsetsPhase Resolution Waveform Size RestrictionsExample Arb Program Program Listing ARB.BASClose System Subroutines Return Print #1, a Send integer value to 4070A Return Return Wiring Diagram DC Operation Option4070A Specifications Operating ModesSoftware CD RemoteHardware Setup Appendix aCLS CLS Print Print Subroutines END SUB Television Remote Control Example Appendix BCLS Make sure a 4070A is attached to the serial port Print Limited Two-Year Warranty Service Information Savi Ranch Pkwy Yorba Linda, CA 92887 USA TEL