B&K 4070A user manual Trigger key, Numeric keys 0 to 9, É, ê, è, ç keys

Page 14

NOTE: Care must be taken when specifying an output offset voltage and level such that the output does not clip. The loaded output cannot swing higher than +6.0V or lower than -6.0V. Therefore:

¦Offset voltage¦ + 1/2 * Vp -p < 6.0

The 4070A can be used as a variable voltage source by setting the output frequency to 0.0 Hz while in Sinewave mode. Then set the offset voltage to the desired output voltage. Remember, the output impedance is 50 ohms. Up to +/- 70 mA can be drawn from the output under these conditions.

This key is also used to select the RS-232 baud rate after the Mode key is pressed once. By pressing Mode then Offset, a menu of baud rates for the 4070A is presented. For more information, refer to the chapter on Remote Mode.

3.5 Trigger key

The Trigger key is used in modes that require a trigger event to begin a process. Triggered Sweep mode uses this key to begin a sweep. Triggered Burst and Data Modulation modes also use this key to begin the generation of a burst signals. Arbitrary Waveform, Function Generator and Pulse Generator modes use this key to begin the generation of a waveform in Triggered mode.

In modes that require a trigger, the trigger can come from three sources:

1.Pressing the Trigger key

2.Applying a low-to-high transition on the Ext Trigger In connector

3.Sending an ASCII "T" to the RS-232 port

The 4070A will simultaneously accept a trigger from all of the above sources.

3.6 Numeric keys (0 to 9, - and.)

The 0 to 9, - and . keys are used to enter a numeric value. The - key is only accepted when entering numeric values that can be negative, i.e. dBm or offset values. The key press is ignored otherwise.

In DTMF generation mode, the. key is used to generate the star ("*") DTMF digit and the - key is used to generate the pound ("#") DTMF digit.

These keys are also used to select a mode after the Mode key is pressed once. These keys are then used to select a mode indicated by the blue text written beneath these keys.

3.7 é, ê, è, ç keys

The arrow keys are used to move the cursor and edit numeric values.

The è and ç keys move the cursor side to side within a cursor field. They can be used to position the cursor over a digit within the field to modify. Once the cursor is over the desired digit, use the é or ê key to increment or decrement that digit.

Pressing the é key increments the digit under the cursor and has the same effect as rotating the knob clockwise 1 tick. Pressing the ê key decrements the digit under the cursor and has the same effect as rotating the knob counter- clockwise 1 tick.

While a numeric value is being typed in, you can use the ç key as an erase key. Pressing this key erases the last digit entered, allowing you to correct typing mistakes.

All arrow keys have an "auto repeat" feature. By holding down the key, the key will repeat continuously until released. This makes it more convenient to quickly sweep a value or move the cursor within a field.

These keys are also used to select a new operating mode after the Mode key is pressed. The arrow keys are then used to select the mode indicated by the blue text beneath the key.

In DTMF Generation mode, these keys specify DTMF digits A, B, C, D which are signaling tones used in the telephone network but not found on a typical telephone. For more information, refer to the chapter on DTMF Generation mode.

9

BK Precision 4070A User Manual Rev.2.2

Image 14
Contents User’s Manual Page Model 4070A User’s Manual +K Precision CorpTable of Contents Arbitrary waveform system Warranty informationRemote operation DC Operation OptionIntroduction Description Feature summary ModesSIG Out connector Installation and setupSync out connector External Modulation In connector External Trigger/Gating/FSK/BPSK input RS232 Interface connector External Reference Input / Ext Arb. Clock connectorExternal Arb Sync In / Out connector Line in connectorField arrow keys Mode keyOperating controls and keys Recall / Store keyÉ, ê, è, ç keys Trigger keyNumeric keys 0 to 9, MHz/dBm, KHz/Vp-p/Sec, Hz/mVp-p/mS keys One Touch Mode KeysClear key Changing frequency Operating mode selectionOperating guide Changing levelEntering a new value Changing valuesModifying an existing value Sinewave Mode 000,000.00 Hz 10.0 dBm Mode descriptionsBasic Sinewave CW Mode Internal AM Mode External AM Input gain 000,000.00 Hz PEP level -10.0 dBm External AM ModeExternal AM Mode Parameters Input GainInt FM 1,000 Hz Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm Internal FM ModeInternal FM Mode Parameters Peak Frequency DeviationExternal FM Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm External FM ModeExternal FM Mode Parameters Peak Phase Deviation Internal PM ModeInternal PM Mode Parameters Int PM 1,000 Hz Pk dev 180 deg 000,000.00 Hz 10.0 dBmExternal PM Pk dev 180 deg 000,000.00 Hz 10.0 dBm External PM ModeExternal PM Mode Parameters Sweep Mode Up / Down sweep Sweep timeMark Frequency Internal FSK ModeInternal FSK Mode Parameters Space FrequencyExt FSK Mark 1,000,000.00 Hz Space 2,000,000.00 Hz 10.0 dBm External FSK ModeExternal FSK Mode Parameters Ext FSK InputBurst frequency Burst Mode IntroductionBurst Mode Trig Ton 10,000 mS 000,000.00 Hz Off 10.0 dBm Int SSB 1,000 Hz Upper Sideband 000,000.00 Hz 10.0 dBm Internal SSB ModeInternal SSB Mode Parameters Upper/Lower Sideband selectionsExternal SSB Upper Sideband 000,000.00 Hz +10.0 dBm External SSB ModeExternal SSB Mode Parameters Dtmf Generation Mode Parameters Dtmf Generation ModeDtmf generation # Duration 200 mS Delay Level -10.0 dBm BackgroundDtmf Delay Dtmf DurationDtmf detection on Ext Mod input 3 # a Dtmf Detection ModeDtmf Detection Mode Parameters Dtmf detected digitsPower Level Power & Voltage Measurement ModePower & Voltage Measurement Mode Parameters Voltage LevelOther mode Arbitrary ModeRemote Mode Int Bpsk 1,000 Hz 000,000.00 Hz 10.0 dBm Internal Bpsk ModeInternal Bpsk Mode Parameters External Bpsk 000,000.00 Hz 10.0 dBm External Bpsk ModeExternal Bpsk Mode Parameters Dualtone Generation Mode 10,000 Hz F2 Ton 000 mS Toff 10.0Dualtone Examples Data Modulation Mode Parameters Data Modulation ModeUsing the Data Modulation Mode Baud RateDownloading Modulation Message Data Example MessageTriggering the Transmission Saving the MessageStart End 1,000,000.00 Hz 10.0 dBm Voltage Controlled Oscillator VCO ModeVCO Mode Parameters End FrequencyChecking your connection with Hyper Terminal ConnectingRemote Operation Remote Control Operation Remote Control Programming Rules× Ø Remote Control CommandsReset 4070A to Sine wave mode Returns CReport hardware and software versions K1.0 Enable, Disable front panel keys and rotary knobE1,0 Enable, Disable LCD echo to terminal ? or H Print help menu Linear Trig Up Time 10,000 mSF0-9 Move cursor to field 0 to Return a CRemote Control Examples ExamplePin 9-pin cable Pin 25 pin cable adapter Arbitrary Waveform ModeQuick Start Guide Introduction to the Arbitrary Waveform System Description of the Arbitrary Waveform GeneratorFunction Generator Switching ModesFeature Summary Pulse GeneratorPhase Arbitrary Waveform ModeInt Clock / Ext Clock / Lock Master Lock slave mode Continuous/Triggered modeWaveform Function Generator ModeFunction Gen Wave Pos Ramp Rep Freq 100,000 Hz 000 mV Repetition FrequencyDuty Cycle Pulse Generator ModePositive Only Saving Arbitrary Waveforms Using WAVELOAD.EXEDownloading arbitrary waveforms Cwaveload /COMx /y /baudz filename.extUsing your own program Rules for Floating Point format Data FormatsFloating Point Format Time & Value Floating Point Format Example #2Digital Format Integer Format Example The data value E468 is handled as follows Hexadecimal FormatRules for Hexadecimal format Example Binary FormatRules for Binary format Generator Multiple Units LockingArb Starting Address = Number of waveform points * Phase Offset More on Phase OffsetsPhase Resolution Waveform Size RestrictionsExample Arb Program Program Listing ARB.BASClose System Subroutines Return Print #1, a Send integer value to 4070A Return Return Wiring Diagram DC Operation Option4070A Specifications Operating ModesSoftware CD RemoteHardware Setup Appendix aCLS CLS Print Print Subroutines END SUB Television Remote Control Example Appendix BCLS Make sure a 4070A is attached to the serial port Print Limited Two-Year Warranty Service Information Savi Ranch Pkwy Yorba Linda, CA 92887 USA TEL