B&K 4070A user manual Multiple Units Locking, Arb, Generator

Page 65

7.9 Multiple Units Locking

7.9.1 Introduction to Multiple Arb Locking

With this feature, users may lock several arbitrary waveform generators together to generate multiple simultaneous signals, which are locked together in frequency. The user may then adjust a phase-offset field to obtain a desired phase relationship between the waveforms with a minimum phase resolution of 0.01 degrees. Any shape of waveform may be used. In fact, waveforms of different shapes may be locked together as long as they all have the same number of points.

Quick Start Guide

To implement Multiple Unit Locking, follow these steps:

1.Cable units together as shown below.

2.Place the Master Unit in Arbitrary Waveform Mode and set the mode on the LCD display to Lock Master

3.Likewise set the mode on each slave unit to Lock Slave

4.Load waveforms into the Master and Slave units. The waveforms may differ in shape but they must each have the same number of points. To generate phase offsets up to 359.99 degrees, the number of points must be either = 32,768 or ≤ 16,384 (refer to the following section “Waveform Size Restrictions”).

5.Set the sampling rate on the Master Unit to the desired sampling frequency. (The frequency shown on the slave units is not used).

6.If desired, set the phase offset on each Slave Unit(s) to the desired phase offset(s).

7.Optionally save the setup on each unit using the Store/Recall button. The waveforms will also be saved.

How it works

Arb wave locking works by using common timing signals for multiple arbitrary waveform generators. To lock multiple generators together, two signals are shared: a sample clock and a reset signal (see the diagram below). The sample clock ensures that the waveform points on all generators are generated at exactly the same rate. The reset signal tells all units to “jump to the first sample point” at the same time. With these two signals, all units are guaranteed to be on the same sample point at the same time. The 4070A uses one arbitrary waveform generator as a “Master” unit that generates the sample clock and reset signal for all of the “Slave” units.

Multi Unit Hookup

Two cables are used to connect on master 4070A to one or more slave units.

NOTE: Observe that one of the connectors for the interconnecting cables is on the Front Panel. You may hook from two to six units together this way.

Master

Slave

Slave

Arb

Arb

Arb

Generator

Generator

Generator

SIG

SYNC

ARB

SIG

Ext Arb

ARB

SIG

Ext Arb

ARB

Out

Out

Sync I/O

Out

Clk In

Sync I/O

Out

Clk In

Sync I/O

(front panel)

(front panel)

(rear panel)

(front panel)

(rear panel)

(rear panel)

(front panel)

(rear panel)

(rear panel)

 

 

Reset Sig

 

 

 

 

 

to

 

 

 

 

 

 

 

 

 

Sample Clock

 

 

 

 

 

more

 

 

 

 

 

 

units

Waveform

Waveform

Waveform

#1

#2

#3

60

BK Precision 4070A User Manual Rev.2.2

Image 65
Contents User’s Manual Page +K Precision Corp Model 4070A User’s ManualTable of Contents Remote operation Warranty informationArbitrary waveform system DC Operation OptionIntroduction Description Modes Feature summarySIG Out connector Installation and setupSync out connector External Modulation In connector External Trigger/Gating/FSK/BPSK input External Arb Sync In / Out connector External Reference Input / Ext Arb. Clock connectorRS232 Interface connector Line in connectorOperating controls and keys Mode keyField arrow keys Recall / Store keyÉ, ê, è, ç keys Trigger keyNumeric keys 0 to 9, MHz/dBm, KHz/Vp-p/Sec, Hz/mVp-p/mS keys One Touch Mode KeysClear key Operating guide Operating mode selectionChanging frequency Changing levelEntering a new value Changing valuesModifying an existing value Sinewave Mode 000,000.00 Hz 10.0 dBm Mode descriptionsBasic Sinewave CW Mode Internal AM Mode External AM Mode Parameters External AM ModeExternal AM Input gain 000,000.00 Hz PEP level -10.0 dBm Input GainInternal FM Mode Parameters Internal FM ModeInt FM 1,000 Hz Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm Peak Frequency DeviationExternal FM Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm External FM ModeExternal FM Mode Parameters Internal PM Mode Parameters Internal PM ModePeak Phase Deviation Int PM 1,000 Hz Pk dev 180 deg 000,000.00 Hz 10.0 dBmExternal PM Pk dev 180 deg 000,000.00 Hz 10.0 dBm External PM ModeExternal PM Mode Parameters Sweep Mode Sweep time Up / Down sweepInternal FSK Mode Parameters Internal FSK ModeMark Frequency Space FrequencyExternal FSK Mode Parameters External FSK ModeExt FSK Mark 1,000,000.00 Hz Space 2,000,000.00 Hz 10.0 dBm Ext FSK InputBurst frequency Burst Mode IntroductionBurst Mode Trig Ton 10,000 mS 000,000.00 Hz Off 10.0 dBm Internal SSB Mode Parameters Internal SSB ModeInt SSB 1,000 Hz Upper Sideband 000,000.00 Hz 10.0 dBm Upper/Lower Sideband selectionsExternal SSB Upper Sideband 000,000.00 Hz +10.0 dBm External SSB ModeExternal SSB Mode Parameters Dtmf generation # Duration 200 mS Delay Level -10.0 dBm Dtmf Generation ModeDtmf Generation Mode Parameters BackgroundDtmf Duration Dtmf DelayDtmf Detection Mode Parameters Dtmf Detection ModeDtmf detection on Ext Mod input 3 # a Dtmf detected digitsPower & Voltage Measurement Mode Parameters Power & Voltage Measurement ModePower Level Voltage LevelOther mode Arbitrary ModeRemote Mode Int Bpsk 1,000 Hz 000,000.00 Hz 10.0 dBm Internal Bpsk ModeInternal Bpsk Mode Parameters External Bpsk 000,000.00 Hz 10.0 dBm External Bpsk ModeExternal Bpsk Mode Parameters 10,000 Hz F2 Ton 000 mS Toff 10.0 Dualtone Generation ModeDualtone Examples Using the Data Modulation Mode Data Modulation ModeData Modulation Mode Parameters Baud RateExample Message Downloading Modulation Message DataSaving the Message Triggering the TransmissionVCO Mode Parameters Voltage Controlled Oscillator VCO ModeStart End 1,000,000.00 Hz 10.0 dBm End FrequencyChecking your connection with Hyper Terminal ConnectingRemote Operation Remote Control Programming Rules Remote Control OperationReset 4070A to Sine wave mode Remote Control Commands× Ø Returns CReport hardware and software versions K1.0 Enable, Disable front panel keys and rotary knobE1,0 Enable, Disable LCD echo to terminal F0-9 Move cursor to field 0 to Linear Trig Up Time 10,000 mS? or H Print help menu Return a CExample Remote Control ExamplesPin 9-pin cable Pin 25 pin cable adapter Arbitrary Waveform ModeQuick Start Guide Description of the Arbitrary Waveform Generator Introduction to the Arbitrary Waveform SystemFeature Summary Switching ModesFunction Generator Pulse GeneratorInt Clock / Ext Clock / Lock Master Lock slave mode Arbitrary Waveform ModePhase Continuous/Triggered modeFunction Gen Wave Pos Ramp Rep Freq 100,000 Hz 000 mV Function Generator ModeWaveform Repetition FrequencyDuty Cycle Pulse Generator ModePositive Only Downloading arbitrary waveforms Using WAVELOAD.EXESaving Arbitrary Waveforms Cwaveload /COMx /y /baudz filename.extUsing your own program Rules for Floating Point format Data FormatsFloating Point Format Example #2 Time & Value Floating Point FormatDigital Format Integer Format Example The data value E468 is handled as follows Hexadecimal FormatRules for Hexadecimal format Example Binary FormatRules for Binary format Generator Multiple Units LockingArb Phase Resolution More on Phase OffsetsStarting Address = Number of waveform points * Phase Offset Waveform Size RestrictionsProgram Listing ARB.BAS Example Arb ProgramClose System Subroutines Return Print #1, a Send integer value to 4070A Return Return DC Operation Option Wiring DiagramOperating Modes 4070A SpecificationsRemote Software CDAppendix a Hardware SetupCLS CLS Print Print Subroutines END SUB Appendix B Television Remote Control ExampleCLS Make sure a 4070A is attached to the serial port Print Limited Two-Year Warranty Service Information Savi Ranch Pkwy Yorba Linda, CA 92887 USA TEL