Agilent Technologies 66311B, D, 66309B, 66111A manual Remote Sense Connections

Page 33

Installation - 3

The maximum allowable value of load lead resistance is 4 ohms total (2 ohms per side). This may be further limited to a lower value, based on peak current loading, by the maximum allowable dc voltage drop of 8 volts total (4 volts per side) as specified for remote sense operation. To illustrate, for up to 2 amps peak, the maximum allowable resistance is 4 ohms total, resulting in a maximum voltage drop of up to 8 volts. For 4 amps peak the maximum allowable resistance is 2 ohms total, again resulting in a maximum allowable voltage drop of up to 8 volts.

In addition to keeping dc resistance low, you also need to minimize the total impedance. For higher slew rate currents (0.2 amps/μs) and long wiring lengths (10 to 20 ft.) the inductance can have as much effect as the resistance. To minimize inductance, twist the load leads. The inductance will be on the order of

0.25μH/ft if twisted, and 0.4 μH/ft if untwisted. In addition to lowering the inductance, twisting the leads will reduce noise pick up. If you are using remote sense leads, connect these as a second twisted pair. Do not twist or bundle them with the load leads.

NOTE: The use of relays between the dc source and the phone also increases impedance. Low resistance relays will improve system performance.

Remote Sense Connections

NOTE: You must use remote sensing on both Output 1 and Output 2 for the unit to operate properly and meet its published specifications. If you are not using output 1 and the open sense protection feature is turned ON, you must jumper the + output 1 pin to its + sense pin, and jumper the - output 1 pin to its - sense pin. Otherwise, the unit will go into a protected state and disable the output (unless open sense protection is turned OFF).

Testing has verified stable performance with up to 20 inches of lead length between the sense lead termination and the phone connection (see figure 3-4). However, for optimum performance, connect the sense leads as close as possible to the phone under test. To minimize inductance, connect the sense leads and load leads as separate twisted pairs (see Figure 3-2).

OUTPUT 1/OUTPUT 2

CONNECTOR

-S - + +S

TWIST LEADS

TWIST PAIR

+

LOAD

_

WIRE RESISTANCE

Figure 3-2. Remote Sense Connections

33

Image 33
Contents USER’S Guide Certification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationDeclaration of Conformity Printing History Acoustic Noise InformationTable of Contents VXIplug&play Power Products Instrument Drivers Checkout Procedure Case of TroubleTypes of Scpi Commands DVM ConnectionsScpi Data Formats Types of Scpi MessagesIntroduction Programming the Output Triggering Output ChangesCalibration Commands 100 Inhibit/Fault IndicatorDisplay Commands 103 Measurement Commands 104Status Commands 121 Output Commands 114System Commands 125 Trigger Commands 126Additional Commands Common CommandsSpecifications 143 Supplemental Characteristics 144 Performing the Calibration Procedure 154Introduction 165 Error Number List 161Basic Introduction 177Front Panel At a Glance Quick ReferenceInstrument Configuration Use the front panel Address key to configure the interfaceRear Panel At a Glance Use the Function keys and Entry keys to enter a new value Front Panel Number EntryImmediate Action Keys Front Panel AnnunciatorsFront Panel Menus At a Glance ABORt CALibrate Scpi Programming Commands At a GlanceUsing the programming interface Using the front panelProgramming the unit using Scpi COMPatibility commands Installing the VXIplug&play instrument driverOptions and Accessories Safety ConsiderationsOption DescriptionAgilent Description and Model DifferencesAgilent 66309B Agilent 66309DCommon Capabilities Front Panel ControlsRemote Programming Dc Source Output 1 Characteristic Output 1 CharacteristicOutput 2 Characteristic Output 2 CharacteristicOption 521 Factory Settings Option 521 Description Agilent 66309B/D onlyOption 521 Relay Modes Page Check the Operating Settings and Conditions Installation and Operation ChecklistAdditional Agilent 66311/66309 Operating Settings Checks Check the Phone ConnectionsInspection CleaningLocation Damage Packaging MaterialInput Connections Connect the Power CordBench Operation Rack MountingOutput Connections Turn the unit off before connecting any wiresOutput Current RatingsRemote Sense Connections Remote Sense ConnectionsRemote Sense Connections with External Relays Maintaining Stability while Remote Sensing Load Regulation and Voltage Drop in the Remote Sense LeadsOpen Sense Lead Protection Message Description Local SensingOutput Compensation OVP Considerations DVM ConnectionsMeasuring Circuits Not Powered by the Main Output Measuring Circuits that are Not Powered by the Main OutputMeasuring Circuits Floating with Respect to the Main Output 10. FLT/INH Examples External Protection ConnectionsDigital I/O Connections FLT/INH DIGital I/O ConnectorComputer Connections Gpib InterfacePin Input/Output Description RS-232 InterfaceProcedure Display Explanation Checkout ProcedurePress Output On/Off Press ProtectProcedure Display Explanation Case of Trouble Selftest Error MessagesPress Shift, Channel Enter Number 12, Enter Press Output On/OffPower-On Selftest Errors Runtime Error MessagesRuntime Error Messages Line FuseFront Panel Description IntroductionFront Panel Operation Display Command Function System KeysOFF Function Keys Immediate Action KeysScrolling Keysq Display Measurement Metering KeysOver Current Output Control KeysEntry Keys Entry KeysExamples of Front Panel Programming Using the Front Panel DisplaySelecting an output on Agilent 66309B/D units Selecting the DVM on Agilent 66311D/66309D unitsSet the output current limit Set the output voltageSet the output compensation Typecap HighSet the output 2 voltage Enable the outputSet the output 2 current limit Keypad, press Enter Number, 7, EnterDisable Overvoltage Protection as follows Querying and Clearing Output Protection and ErrorsQuery and Clear Errors as follows OvercurrentUse the Meter menu for making front panel measurements Making Basic Front Panel MeasurementsMaking Enhanced Front Panel Measurements Default Front Panel Measurement ParametersUse the Meter menu for making DVM measurements Making DVM Measurements Agilent 66311D/66309D onlyCurrrang Auto Currdet AcdcTo configure the Ridfi mode of the port, proceed as follows Setting the Gpib Address and Programming LanguageTo configure the Digio mode of the port, proceed as follows Set the Gpib address as followsStoring and Recalling Instrument States External References Gpib ReferencesScpi References Downloading and Installing the Driver VXIplug&play Power Products Instrument DriversSupported Applications System RequirementsGpib Capabilities of the DC Source Accessing Online HelpRS-232 Capabilities of the DC Source Gpib AddressIntroduction to Scpi RS-232 Flow ControlBaud Rate Types of Scpi Commands Conventions Used in This GuideBoldface font Multiple Commands in a Message Including Common CommandsMoving Among Subsystems Types of Scpi Messages Using QueriesMessage Unit HeadersScpi Data Formats Suffixes and Multipliers Scpi Command CompletionResponse Data Types Class Suffix Unit Unit with MultiplierScpi Conformed Commands Using Device ClearNon-SCPI Commands Scpi Conformance InformationPower-on Initialization Programming the OutputEnabling the Output Output Current Output VoltageTriggering Output Changes Setting the Voltage or Current Transient LevelsScpi Triggering Nomenclature Output Trigger ModelSelecting the Output Trigger Source Enabling the Output Trigger SystemGenerating Triggers Single TriggerMaking Basic Measurements Average MeasurementsControlling Measurement Samples Window Functions Making Enhanced MeasurementsRMS Measurements Current Ranges and Measurement DetectorPulse Measurements Minimum and Maximum MeasurementsHigh/Low Measurements Returning All Measurement Data From the Data Buffer Making DVM MeasurementsTriggered Measurements Measurement Trigger ModelSequence Form Alias SEQuence2 ACQuire Selecting the Measurement Trigger Source Enabling the Measurement Trigger SystemSelecting the Sensing Function INTernalSingle Triggers Generating Measurement TriggersTrigacqcouncurr number or Trigacqcounvolt number Pre-trigger and Post-trigger Data Acquisition Programming the Status RegistersDC Source Status Model Power-On ConditionsOperation Status Group Bit Configurations of Status RegistersQuestionable Status Group PON Power On BitStandard Event Status Group Status Byte RegisterServicing Operation Status and Questionable Status Events Determining the Cause of a Service InterruptMSS Bit RQS BitInhibit/Fault Indicator Monitoring Both Phases of a Status TransitionRemote Inhibit RI Using the Inhibit/Fault Port as a Digital I/O Discrete Fault Indicator DFIBit Weight PinSubsystem Commands Subsystem Commands Syntax Language Dictionary Language Dictionary Programming Parameters Common CommandsCALibrateCURRent Calibration CommandsCALibrateCURRent2 CALibrateCURRentMEASureLOWRangeCALibrateDATA CALibratePASSwordCALibrateDATE CALibrateDVMCALibrateSTATe CALibrateSAVECALibrateVOLTage CALibrateVOLTage2DISPlay Display CommandsDISPlayCHANnel DISPlayMODEMeasurement Commands FORMat104 MEASureARRayCURRent? FETChARRayCURRent? FORMatBORDerMEASureARRayVOLTage? FETChARRayVOLTage? Query SyntaxMEASureCURRent2? MEASureCURRent? FETChCURRent?MEASureCURRentACDC? FETChCURRentACDC? NR3MEASureCURRentLOW? FETChCURRentLOW? MEASureCURRentHIGH? FETChCURRentHIGH?MEASureCURRentMAXimum? FETChCURRent MAXimum? 107MEASureDVM? FETChDVM? MEASureCURRentMINimum? FETChCURRentMINimum?MEASureDVMACDC? FETChDVMACDC? MEASureVOLTage? FETChVOLTage?MEASureVOLTageACDC? FETChVOLTageACDC? MEASureVOLTage2MEASureVOLTageHIGH? FETChVOLTageHIGH? 109MEASureVOLTageMAXimum? FETChVOLTageMAXimum? MEASureVOLTageLOW? FETChVOLTageLOW?MEASureVOLTageMINimum? FETChVOLTageMINimum? 110SENSeCURRentRANGe SENSeCURRentDETectorQuery Syntax SENSeCURRentDETector? 111SENSePROTectionSTATe SENSeFUNCtionSENSeSWEepOFFSetPOINts SENSeSWEepPOINtsSENSeWINDow SENSeSWEepTINTervalQuery Syntax SENSeSWEepTINTerval? Query Syntax SENSeWINDowTYPE?INSTrumentCOUPleOUTPutSTATe Output CommandsOUTPut1 OUTPut1 2RELayMODEOUTPutDFISOURce OUTPutDFIOUTPutPONSTATe 115OUTPutPROTectionCLEar OUTPutPROTectionDELayOUTPutRIMODE OUTPutTYPESOURceCURRent SOURceCURRent2117 SOURceCURRentPROTectionSTATe Command Syntax Parameters RST Value ExamplesSOURceCURRentTRIGger SOURceCURRent2TRIGgerSOURceDIGitalFUNCtion SOURceDIGitalDATASOURceVOLTage 119SOURceVOLTagePROTection SOURceVOLTage2SOURceVOLTagePROTectionSTATe 120STATusPRESet Status CommandsSOURceVOLTageTRIGger SOURceVOLTage2TRIGgerBit Configuration of Operation Status Registers STATusOPERationENABleSTATusOPERation? STATusOPERationCONDition?Bit Configuration of Questionable Status Registers Parameters Preset ValueSTATusOPERationNTR STATusOPERationPTR STATusQUEStionable?STATusQUEStionableCONDition? STATusQUEStionableENABleSTATusQUEStionableNTR STATusQUEStionablePTR 124SYSTemERRor? System CommandsSYSTemLANGuage SYSTemVERSion?Related Commands Trigger CommandsABORt INITiateSEQuence INITiateNAMETRIGger TRIGgerSOURce127 TRIGgerSEQuence2COUNtCURRent TRIGgerACQuireCOUNtCURRent TRIGgerSEQuence2 TRIGgerACQuireTRIGgerSEQuence2COUNtDVM TRIGgerACQuireCOUNtDVM 128TRIGgerSEQuence2COUNtVOLTage TRIGgerACQuireCOUNtVOLTage Unit129 130 TRIGgerSEQuence2HYSTeresisDVM TRIGgerACQuireHYSTeresisDVMTRIGgerSEQuence2LEVelCURRent TRIGgerACQuireLEVelCURRent TRIGgerSEQuence2LEVelDVM TRIGgerACQuireLEVelDVM131 TRIGgerSEQuence2SLOPeCURRent TRIGgerACQuireSLOPeCURRent TRIGgerSEQuence2LEVelVOLTage TRIGgerACQuireLEVelVOLTageReturned Parameters 132TRIGgerSEQuence2SLOPeDVM TRIGgerACQuireSLOPeDVM TRIGgerSEQuence2SLOPeVOLTage TRIGgerACQuireSLOPeVOLTage133 TRIGgerSEQuence2SOURce TRIGgerACQuireSOURce TRIGgerSEQuence1DEFine TRIGgerSEQuence2DEFine134 Command Syntax *CLS Parameters None Common CommandsBit Configuration of Standard Event Status Enable Register 135Field Information Example136 RCL NRf Query Syntax *OPT? Returned Parameters AardParameters Example 137Command Syntax *SAV NRf Example *SAV Related Commands *RCL *RST 138RST Settings Bit Configuration of Status Byte Register Power-on ValueQuery Syntax *STB? 139Query Syntax TST? Returned Parameters NR1 TRGAdditional Commands VOLTageLIMitHIGH? CURRentPROTectionTRIPped?VOLTageLIMitLOW? VOLTagePROTectionTRIPped?Specifications Table A-1. Performance Specifications143 Table A-2. Supplemental Characteristics Supplemental Characteristics144 Agilent 66309B/D145 146 Table A-4. Agilent 66309B/D Option 521 CharacteristicsEquipment Required Test Setup147 Table B-1. Equipment RequiredPerforming the Verification Tests Turn-On Checkout148 Current Programming and Measurement Accuracy Voltage Programming and Measurement AccuracyOutput terminals on both outputs 1 Steps 6-10 apply to Agilent 66309B/D output 2 onlySteps 7-11 apply to Agilent 66309B/D output 2 only Current Sink Measurement150 151 DVM Measurement Accuracy152 DVM Voltage Measurement Agilent 66309D only 153Performing the Calibration Procedure Front Panel Calibration Menu154 Check the Language Setting Front Panel Calibration ProcedureEnable Calibration Mode Voltage Programming and Measurement CalibrationOvervoltage Protection Calibration Steps 11-16 apply to Agilent 66309B/D output 2 onlyCurrent Programming and High-Range Measurement Calibration 156Steps 27-32 apply to Agilent 66309B/D output 2 only 157Calcurrmeas AC DVM Calibration applies to Agilent 66311D, 66309D only Restore the Language SettingSaving the Calibration Constants 158Changing the Calibration Password Calibration Error MessagesTable B-3. Gpib Calibration Error Messages Calibration Over the GpibPage Error Number List Table C-1. Error Numbers161 162 Error Messages163 Error Messages CPage Assigning the Gpib Address in Programs National Instruments Gpib Driver165 Error Handling Example 1. National Instruments Interface Example166 Example Programs D 167168 Example 2. Controller Using Basic169 Example 3. Current Pulse Measurement Using Basic170 171 172 Example 5. DFI Example Using Basic DFI Programming Example173 Page Install the Correct Line Fuse Configure the Power TransformerOpen the Unit Close the Unit176 Figure E-1, Power Transformer AC Input ConnectionsCommand Setting Table F-1. COMPatibility Power-on Settings177 Similar Scpi Command Compatibility CommandTable F-2. COMPatibility Commands 178179 NumberError String Description/Explanation/Examples ErrorTable F-3. COMPatibility Errors 180181 Table F-5. Bit Configuration of Serial Poll RegisterPage 183 Index184 Index185 186 OVERTEMPERATURE, 47 OVERVOLTAGE, 47 OVLD, 47, 60, 61187 Subsystem commands syntax, 96 suffixesType CAP 188Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates
Related manuals
Manual 8 pages 56.17 Kb

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.