Agilent Technologies 66309B STATusOPERation?, STATusOPERationCONDition?, STATusOPERationENABle

Page 122

8 – Language Dictionary

 

 

 

 

 

 

 

 

Table 8-4. Bit Configuration of Operation Status Registers

 

Bit Position

15–13

12

11

10

9

8

7-6

5

4-1

Bit Name

not

CC2

CC-

CC+

CV2

CV

not

WTG

not

 

used

 

 

 

 

 

used

 

used

Bit Weight

 

4096

2048

1024

512

256

 

32

 

0

CAL

1

CAL = The dc source is computing new calibration constants. WTG = The dc source is waiting for a trigger.

CV = The dc source is operating in constant voltage mode.

CV2 = Output 2 is operating in constant voltage mode. (Agilent 66309B/D only) CC+ = The dc source is operating in constant current mode.

CC- = The dc source is operating in negative constant current mode.

CC2 = Output 2 is operating in constant current mode. (Agilent 66309B/D only)

STATus:OPERation?

This query returns the value of the Operation Event register. The Event register is a read-only register, which holds (latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation Event register clears it.

Query Syntax

STATus:OPERtion[:EVENt]?

Parameters

None

Returned Parameters

<NR1>(Register Value)

Examples

STAT:OPER?

Related Commands

*CLS STAT:OPER:NTR STAT:OPER:PTR

STATus:OPERation:CONDition?

This query returns the value of the Operation Condition register. That is a read-only register, which holds the real-time (unlatched) operational status of the dc source.

Query Syntax

STATus:OPERation:CONDition?

Parameters

None

Examples

STAT:OPER:COND? STATUS:OPERATION:CONDITION?

Returned Parameters

<NR1> (register value)

STATus:OPERation:ENABle

This command and its query set and read the value of the Operational Enable register. This register is a mask for enabling specific bits from the Operation Event register to set the operation summary bit (OPER) of the Status Byte register. This bit (bit 7) is the logical OR of all the Operatonal Event register bits that are enabled by the Status Operation Enable register.

Command Syntax

STATus:OPERation:ENABle<NRf>

Parameters

0 to 32767

Preset Value

0

Examples

STAT:OPER:ENAB 1312

Query Syntax

STATus:OPERation:ENABle?

Returned Parameters

<NR1> (register value)

Related Commands

STAT:OPER:EVEN

122

Image 122
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration EMCDeclaration of Conformity Acoustic Noise Information Printing HistoryTable of Contents Types of Scpi Commands Checkout Procedure Case of TroubleVXIplug&play Power Products Instrument Drivers DVM ConnectionsIntroduction Programming the Output Types of Scpi MessagesScpi Data Formats Triggering Output ChangesDisplay Commands 103 Inhibit/Fault IndicatorCalibration Commands 100 Measurement Commands 104System Commands 125 Output Commands 114Status Commands 121 Trigger Commands 126Specifications 143 Supplemental Characteristics 144 Common CommandsAdditional Commands Performing the Calibration Procedure 154Basic Error Number List 161Introduction 165 Introduction 177Quick Reference Front Panel At a GlanceRear Panel At a Glance Instrument ConfigurationUse the front panel Address key to configure the interface Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Scpi Programming Commands At a Glance ABORt CALibrateProgramming the unit using Scpi COMPatibility commands Using the front panelUsing the programming interface Installing the VXIplug&play instrument driverOption Safety ConsiderationsOptions and Accessories DescriptionAgilent 66309B Description and Model DifferencesAgilent Agilent 66309DRemote Programming Common CapabilitiesFront Panel Controls Output 1 Characteristic Dc Source Output 1 CharacteristicOutput 2 Characteristic Output 2 CharacteristicOption 521 Relay Modes Option 521 Factory SettingsOption 521 Description Agilent 66309B/D only Page Additional Agilent 66311/66309 Operating Settings Checks Installation and Operation ChecklistCheck the Operating Settings and Conditions Check the Phone ConnectionsLocation CleaningInspection Damage Packaging MaterialBench Operation Connect the Power CordInput Connections Rack MountingOutput Turn the unit off before connecting any wiresOutput Connections Current RatingsRemote Sense Connections Remote Sense ConnectionsRemote Sense Connections with External Relays Load Regulation and Voltage Drop in the Remote Sense Leads Maintaining Stability while Remote SensingOpen Sense Lead Protection Local Sensing Message DescriptionOutput Compensation DVM Connections OVP ConsiderationsMeasuring Circuits that are Not Powered by the Main Output Measuring Circuits Not Powered by the Main OutputMeasuring Circuits Floating with Respect to the Main Output External Protection Connections 10. FLT/INH ExamplesComputer Connections FLT/INH DIGital I/O ConnectorDigital I/O Connections Gpib InterfaceRS-232 Interface Pin Input/Output DescriptionPress Output On/Off Checkout ProcedureProcedure Display Explanation Press ProtectProcedure Display Explanation Press Shift, Channel Selftest Error MessagesCase of Trouble Enter Number 12, Enter Press Output On/OffRuntime Error Messages Runtime Error MessagesPower-On Selftest Errors Line FuseIntroduction Front Panel DescriptionFront Panel Operation OFF Display Command FunctionSystem Keys Scrolling Keysq Function KeysImmediate Action Keys Metering Keys Display MeasurementOutput Control Keys Over CurrentEntry Keys Entry KeysSelecting an output on Agilent 66309B/D units Using the Front Panel DisplayExamples of Front Panel Programming Selecting the DVM on Agilent 66311D/66309D unitsSet the output compensation Set the output voltageSet the output current limit Typecap HighSet the output 2 current limit Enable the outputSet the output 2 voltage Keypad, press Enter Number, 7, EnterQuery and Clear Errors as follows Querying and Clearing Output Protection and ErrorsDisable Overvoltage Protection as follows OvercurrentMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsDefault Front Panel Measurement Parameters Making Enhanced Front Panel MeasurementsCurrrang Auto Making DVM Measurements Agilent 66311D/66309D onlyUse the Meter menu for making DVM measurements Currdet AcdcTo configure the Digio mode of the port, proceed as follows Setting the Gpib Address and Programming LanguageTo configure the Ridfi mode of the port, proceed as follows Set the Gpib address as followsStoring and Recalling Instrument States Scpi References External ReferencesGpib References Supported Applications VXIplug&play Power Products Instrument DriversDownloading and Installing the Driver System RequirementsRS-232 Capabilities of the DC Source Accessing Online HelpGpib Capabilities of the DC Source Gpib AddressBaud Rate Introduction to ScpiRS-232 Flow Control Boldface font Types of Scpi CommandsConventions Used in This Guide Moving Among Subsystems Multiple Commands in a MessageIncluding Common Commands Message Unit Using QueriesTypes of Scpi Messages HeadersScpi Data Formats Response Data Types Scpi Command CompletionSuffixes and Multipliers Class Suffix Unit Unit with MultiplierNon-SCPI Commands Using Device ClearScpi Conformed Commands Scpi Conformance InformationEnabling the Output Power-on InitializationProgramming the Output Output Voltage Output CurrentScpi Triggering Nomenclature Setting the Voltage or Current Transient LevelsTriggering Output Changes Output Trigger ModelGenerating Triggers Enabling the Output Trigger SystemSelecting the Output Trigger Source Single TriggerControlling Measurement Samples Making Basic MeasurementsAverage Measurements Making Enhanced Measurements Window FunctionsCurrent Ranges and Measurement Detector RMS MeasurementsHigh/Low Measurements Pulse MeasurementsMinimum and Maximum Measurements Making DVM Measurements Returning All Measurement Data From the Data BufferSequence Form Alias SEQuence2 ACQuire Triggered MeasurementsMeasurement Trigger Model Selecting the Sensing Function Enabling the Measurement Trigger SystemSelecting the Measurement Trigger Source INTernalGenerating Measurement Triggers Single TriggersTrigacqcouncurr number or Trigacqcounvolt number Programming the Status Registers Pre-trigger and Post-trigger Data AcquisitionPower-On Conditions DC Source Status ModelBit Configurations of Status Registers Operation Status GroupStandard Event Status Group PON Power On BitQuestionable Status Group Status Byte RegisterMSS Bit Determining the Cause of a Service InterruptServicing Operation Status and Questionable Status Events RQS BitRemote Inhibit RI Inhibit/Fault IndicatorMonitoring Both Phases of a Status Transition Bit Weight Discrete Fault Indicator DFIUsing the Inhibit/Fault Port as a Digital I/O PinSubsystem Commands Subsystem Commands Syntax Language Dictionary Language Dictionary Common Commands Programming ParametersCALibrateCURRent2 Calibration CommandsCALibrateCURRent CALibrateCURRentMEASureLOWRangeCALibrateDATE CALibratePASSwordCALibrateDATA CALibrateDVMCALibrateVOLTage CALibrateSAVECALibrateSTATe CALibrateVOLTage2DISPlayCHANnel Display CommandsDISPlay DISPlayMODE104 Measurement CommandsFORMat MEASureARRayVOLTage? FETChARRayVOLTage? FORMatBORDerMEASureARRayCURRent? FETChARRayCURRent? Query SyntaxMEASureCURRentACDC? FETChCURRentACDC? MEASureCURRent? FETChCURRent?MEASureCURRent2? NR3MEASureCURRentMAXimum? FETChCURRent MAXimum? MEASureCURRentHIGH? FETChCURRentHIGH?MEASureCURRentLOW? FETChCURRentLOW? 107MEASureDVMACDC? FETChDVMACDC? MEASureCURRentMINimum? FETChCURRentMINimum?MEASureDVM? FETChDVM? MEASureVOLTage? FETChVOLTage?MEASureVOLTageHIGH? FETChVOLTageHIGH? MEASureVOLTage2MEASureVOLTageACDC? FETChVOLTageACDC? 109MEASureVOLTageMINimum? FETChVOLTageMINimum? MEASureVOLTageLOW? FETChVOLTageLOW?MEASureVOLTageMAXimum? FETChVOLTageMAXimum? 110Query Syntax SENSeCURRentDETector? SENSeCURRentDETectorSENSeCURRentRANGe 111SENSeSWEepOFFSetPOINts SENSeFUNCtionSENSePROTectionSTATe SENSeSWEepPOINtsQuery Syntax SENSeSWEepTINTerval? SENSeSWEepTINTervalSENSeWINDow Query Syntax SENSeWINDowTYPE?OUTPut1 Output CommandsINSTrumentCOUPleOUTPutSTATe OUTPut1 2RELayMODEOUTPutPONSTATe OUTPutDFIOUTPutDFISOURce 115OUTPutRIMODE OUTPutPROTectionDELayOUTPutPROTectionCLEar OUTPutTYPE117 SOURceCURRentSOURceCURRent2 SOURceCURRentTRIGger Command Syntax Parameters RST Value ExamplesSOURceCURRentPROTectionSTATe SOURceCURRent2TRIGgerSOURceVOLTage SOURceDIGitalDATASOURceDIGitalFUNCtion 119SOURceVOLTagePROTectionSTATe SOURceVOLTage2SOURceVOLTagePROTection 120SOURceVOLTageTRIGger Status CommandsSTATusPRESet SOURceVOLTage2TRIGgerSTATusOPERation? STATusOPERationENABleBit Configuration of Operation Status Registers STATusOPERationCONDition?STATusOPERationNTR STATusOPERationPTR Parameters Preset ValueBit Configuration of Questionable Status Registers STATusQUEStionable?STATusQUEStionableNTR STATusQUEStionablePTR STATusQUEStionableENABleSTATusQUEStionableCONDition? 124SYSTemLANGuage System CommandsSYSTemERRor? SYSTemVERSion?ABORt Trigger CommandsRelated Commands INITiateSEQuence INITiateNAME127 TRIGgerTRIGgerSOURce TRIGgerSEQuence2COUNtDVM TRIGgerACQuireCOUNtDVM TRIGgerSEQuence2 TRIGgerACQuireTRIGgerSEQuence2COUNtCURRent TRIGgerACQuireCOUNtCURRent 128129 TRIGgerSEQuence2COUNtVOLTage TRIGgerACQuireCOUNtVOLTageUnit TRIGgerSEQuence2HYSTeresisDVM TRIGgerACQuireHYSTeresisDVM 130131 TRIGgerSEQuence2LEVelCURRent TRIGgerACQuireLEVelCURRentTRIGgerSEQuence2LEVelDVM TRIGgerACQuireLEVelDVM Returned Parameters TRIGgerSEQuence2LEVelVOLTage TRIGgerACQuireLEVelVOLTageTRIGgerSEQuence2SLOPeCURRent TRIGgerACQuireSLOPeCURRent 132133 TRIGgerSEQuence2SLOPeDVM TRIGgerACQuireSLOPeDVMTRIGgerSEQuence2SLOPeVOLTage TRIGgerACQuireSLOPeVOLTage 134 TRIGgerSEQuence2SOURce TRIGgerACQuireSOURceTRIGgerSEQuence1DEFine TRIGgerSEQuence2DEFine Bit Configuration of Standard Event Status Enable Register Common CommandsCommand Syntax *CLS Parameters None 135136 Field InformationExample Parameters Example Query Syntax *OPT? Returned Parameters AardRCL NRf 137RST Settings Command Syntax *SAV NRfExample *SAV Related Commands *RCL *RST 138 Query Syntax *STB? Power-on ValueBit Configuration of Status Byte Register 139TRG Query Syntax TST? Returned Parameters NR1Additional Commands VOLTageLIMitLOW? CURRentPROTectionTRIPped?VOLTageLIMitHIGH? VOLTagePROTectionTRIPped?143 SpecificationsTable A-1. Performance Specifications 144 Supplemental CharacteristicsTable A-2. Supplemental Characteristics Agilent 66309B/D145 Table A-4. Agilent 66309B/D Option 521 Characteristics 146147 Test SetupEquipment Required Table B-1. Equipment Required148 Performing the Verification TestsTurn-On Checkout Output terminals on both outputs 1 Voltage Programming and Measurement AccuracyCurrent Programming and Measurement Accuracy Steps 6-10 apply to Agilent 66309B/D output 2 only150 Steps 7-11 apply to Agilent 66309B/D output 2 onlyCurrent Sink Measurement DVM Measurement Accuracy 151152 153 DVM Voltage Measurement Agilent 66309D only154 Performing the Calibration ProcedureFront Panel Calibration Menu Enable Calibration Mode Front Panel Calibration ProcedureCheck the Language Setting Voltage Programming and Measurement CalibrationCurrent Programming and High-Range Measurement Calibration Steps 11-16 apply to Agilent 66309B/D output 2 onlyOvervoltage Protection Calibration 156Calcurrmeas AC Steps 27-32 apply to Agilent 66309B/D output 2 only157 Saving the Calibration Constants Restore the Language SettingDVM Calibration applies to Agilent 66311D, 66309D only 158Table B-3. Gpib Calibration Error Messages Calibration Error MessagesChanging the Calibration Password Calibration Over the GpibPage 161 Error Number ListTable C-1. Error Numbers Error Messages 162Error Messages C 163Page 165 Assigning the Gpib Address in ProgramsNational Instruments Gpib Driver 166 Error HandlingExample 1. National Instruments Interface Example 167 Example Programs DExample 2. Controller Using Basic 168Example 3. Current Pulse Measurement Using Basic 169170 171 172 173 Example 5. DFI Example Using BasicDFI Programming Example Page Open the Unit Configure the Power TransformerInstall the Correct Line Fuse Close the UnitFigure E-1, Power Transformer AC Input Connections 176177 Command SettingTable F-1. COMPatibility Power-on Settings Table F-2. COMPatibility Commands Compatibility CommandSimilar Scpi Command 178179 Table F-3. COMPatibility Errors ErrorNumberError String Description/Explanation/Examples 180Table F-5. Bit Configuration of Serial Poll Register 181Page Index 183Index 184185 OVERTEMPERATURE, 47 OVERVOLTAGE, 47 OVLD, 47, 60, 61 186Subsystem commands syntax, 96 suffixes 187188 Type CAPEurope Asia Pacific United States Latin AmericaCanada Australia/New Zealand JapanManual Updates
Related manuals
Manual 8 pages 56.17 Kb

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.